Skip to main content

Full text of "USPTO Patents Application 10687850"

See other formats


per 



WORLD INTELLECTUAL PROPERTY ORGANIZATION 
International Bureau 




INTERNATIONAL APPLICATION PUBLISHED UNDER THE PATENT COOPERATION TREATY (P CI) 

(1 1) International Publication Number: WO 92/01928 

(43) International Publication Date: 6 February 1992 (06.02.92) 



(51) International Patent aassification 5 
G01N 27/30, 31/12, 27/02 
G01N 27/416, 27/28, 27/00 
COIN 31/00, G01F1/66 



Al 



(21) International Application Number: 



PCT/US91/0507I 



(22) InterarSonal Filing Date: 18 July 1991 (18.07.91) 

20 July 1990 (20.07.90) US 



(30) Priority data 
555,976 



(71) Applicant: I-STAT CORPORATION [US/US]; 303 Col- 
lege Road East, Princeton, NJ 08540 (US). 

(72) Inventors: COZZETTE, Stephen, N. ; 45 Dennis Court, 
Hjghtown, NJ 08520 (US). DAVIS, Graham ; 15-04 Fox 
Run Drive, Plainsboro, NJ 08536 (US). HOLLERIT- 
mS* 7^T^ n i 14 Sunset Unt > 0ak Wd » NJ 07438 
£3. R ' ; 10,1 Yardley-MorrisviHe 
Road, Yardley, PA 19067 (US). PIZNIK, Sylvia ; 12 

, nn o.o owt » Jackson > N J 08527 (US). SMIT, Nicola- 
llin fh^^Yr^\ tn ^ Woodkwn. Ontario K0A 
3M0 (CA). TIRINATO, Jody, Ann ; 27-06 Hunters Glen 

?i r S^ P,a, n nsb S r S!r 08536 < US >- mEC *> Henry, J. ; 
31 Parker Road, Piamsboro, NJ 08536 (US). ZELIN, Mil 
chad, P. ; 9104 Tamarrbn Drive, Plainsboro, NJ 08536 
(US). 



(74) Agent: MISROCK, S., Lesbe; Pennie & Edmonds, 1155 
Avenue of the Americas, New York, NY 10036 (US). 

ffi C ^ CH (European patent), DE (EuropeanV 
ent), DK (European patent), ES (European patent), FR 
^uropean patent) GB (European patent), GR (Euro- 

pean patent), NL (European patent), SE (European pa- 

Published 

With international search report 
Before the expiration of the time limit for amending the 
claims and to be republished in the event of the receipt of\ 
amendments, 1 



' (54)TMe: METHOD FOR ANALYTICALLY UTILIZING MICROFABRICATED SENSORS DURING WET-UP 

(57) At act 

first time window in the presence of the i«S KZ„1^ ^L?k $ Pfrf»nnu,g a first signal measurement in a 
ment; (I) contacting the ^SS^SS^SS^^t ftid 1 T ? e first signal measure - 

ond time window in the presence of the^nleS a^M ^ performing a second s.gnal measurement in a sec 



FOR THE PURPOSES OF INFORMATION ONLY 



Codes used Co identify States party to the PCT on the front pages of pamphlets publishing international 
applications under the PCT. 



AT 


Austria 


ES 


Spain 


MG 


Madagascar 


AU 


Australia 


Fl 


Finland 


ML 


Mali 


an 


Barbados 


PR 


France 


MN 


Mongolia 


BB 


Belgium 


CA 


Gabon 




Mauritania 


BF 


Burkina Faso 


CB 


United Kingdom 


MW 


Malawi 


BG 


Bulgaria 


CN 


Guinea 


NL 


Netherlands 


BJ 


Bcnio 


CR 


Greece 


NO 


Norway 


BR 


Brazil 


HU 


Hungary 


PL 


Poland 


CA 


Canada 


IT 


Italy 


RO 


Romania 


CP 


Central African Republic 


JP 


Japan 


SD 


Sudan 


CG 


Congo 


KP 


Democratic People's Republic 


SB 


Sweden 


CH 


Switzerland 




or Korea 


SN 


Senegal 


a 


CStc dlvoire 


KR 


Republic of Korea 


su* 


Soviet Union 


CM 


Cameroon 


Lf 


1 ittfcfrtcnstcin 


TD 


Chad 


CS 


Czechoslovakia 


LK 


Sri Lanka 


TC 


Togo 


DE 


Germany 


LU 




US 


United States of America 


DK 


Denmark 


MC 


Monaco 







+ It is not yet known for which States of the former Soviet Union any designation of the 
Soviet Union has effect. 



WO 92/01928 



-1- 



PCIYUS91/05071 



METHOD FOR ANALYTICALLY UTILIZING 
MI CROFABRI CATED SENSORS DURING WET-UP 



CROSS-REFERENCE TO RELATRn APPLTPATTOWg 

5 The present U. S. Application is related to 

prior co-pending U. S. Application Serial Nos. 07/432,714, 
filed November 7, 1989, and 07/245,102, filed September 15, 
1988, the disclosures of which are incorporated by 
reference herein in their entirety, 

10 1 . FIELD OF THE INVENTION 

The present invention relates to a method of 
quantifying a preselected analyte species present in 
fluids, which takes advantage of the well-behaved 
equilibration wet-up characteristics of dry-stored wholly 

15 microfabricated electrochemical sensors. These and other 
performance characteristics obtain from a manufacturing 
process, described in the co-pending U.S. Application 
Serial No.. 07/432,714, that attains a high degree of 
uniformity with regard to the physical dimensions and 

20 resulting ; properties of such electrochemical sensors. The 
invention allows for the acquisition of analytical data 
while the signal of the sensor of interest is still 
undergoing the process of an equilibration wet-up. In 
particular, the present method includes deriving useful 

25 information from microfabricated electrochemical sensors, 
which sensors had been stored dry, much more quickly than 
previously thought practical by acquiring and manipulating 
selected signal measurements well before the sensors have 
attained a post-equilibrated wet-up state (i.e., steady- 

30 state) response. Most notably, the invention provides a 
method for relating the signal measurements recorded in 
different fluids to determine the ratio of the 



WO 92/01928 



- 2 - 



PCT/US91/05071 



concentration of a preselected analyte species in each 
fluid. The present method utilizes a computational means 
which is able to distinguish the signal response of a given 
electrochemical sensor to changes in analyte concentration, 
5 which response is fast relative to the slower monotonic 
"wet-up" behavior of the sensor and its~ associated 
reference electrode. 

2 . BACKGROUND OF THE INVENTION 

The recent emphasis in the development of 

10 clinical chemistry technology has been directed toward the 
development of systems for "real time" analysis of 
biological fluids or those analyses which can be performed 
in the close proximity of the patient e.g., at the bedside 
or in the physician's office. Such biological fluids 

15 include urine, plasma, serum, and preferably, whole-blood. 
Clear benefits are . achieved if the chemical information 
required by the physician is obtained during patient 
consultation and not several hours or days afterward. 
Although progress has been made toward achieving such a 

20 goal, many problems still remain including the limitations 
of established manufacturing methods to mass-produce 
electrochemical devices with sufficiently uniform 
performance characteristics and extended shelf-lives. Of 
particular interest is the lack of adequate computational 

25 techniques which minimize the time required to obtain 
useful information from such electrochemical devices. 

To date, fluid analysis has been carried out 
using many types of electrochemical sensors in which 
potentiometric, amperometric or conduct imetric 
30 measurements are performed in a steady-state or kinetic 
(e.g., initial rate) mode. Electrochemical sensors 
employed for these measurements usually consist of e two- 
component assembly in which a sensitized membrane is 
interposed between the fluid and an underlying electrode. 



WO 92/01928 



- 3 - 



PCT/US91/05071 



Some membrane compositions have distinctive species 
recognition capabilities which enable the electrochemical 
sensor to detect the analyte of interest specifically and 
measure its concentration in a complex biological fluid. 
5 To date, however, size, complexity and expense limitations, 
combined with a high incidence of error resulting both from 
instrument quality control and accidental errors by the 
operator, have impeded wide spread use of this technology 
in locations, such as the emergency room and the doctor's 

10 office, which are remote from the central clinical 
chemistry laboratory. Moreover, most analytical test 
methods currently in use are overly cumbersome or complex. 
More significantly, the response of the electrochemical 
device, itself, may be so slow as to make such "real time" 

15 analysis very difficult. It should be noted that the 
concentration of certain components, such as glucose and 
potassium ion, in a biological fluid (e.g., whole-blood) 
may change significantly over a prolonged period. The 
change arises likely from hemolysis "of related metabolic 

20 processes. 

As mentioned previously, a principal obstacle 
against the successful implementation of "real time" 
clinical fluid analysis is the lack of reliable sensor 
manufacturing methods. Equally lacking, however, are data 
25 acquisition and manipulation methods which allow the quick 
retrieval of information from existing chemical sensing 
devices some of which are stored substantially dry in order 
to maximize shelf-life. The prevailing standard practice 
dictates that these "dry-stored" devices be allowed to 
reach a fully equilibrated "wet-up" state before meaningful 
sensor data can be recorded. 



30 



WO 92/01928 PCT/US91/05071 

- 4 - 



2.1. PREVIOUS DEVICES AND METHODS 
FOR FLUID ANALYSIS 

Some progress has been made toward the 
production of improved testing apparatus, including 
5 miniaturized chemical sensing devices. U. S. Patent No. 
4,734,184 issued to Burleigh et al. discloses an electrode 
assembly for monitoring the concentration of a number of 
gases and ions present in the blood. Although the assembly 
is stored dry to promote an extended shelf-life, the 

10 electrodes, are thoroughly hydrated (wet-up) prior to use. 
During operation they are in prolonged and equilibrated 
contact with the many solutions employed/ including a 
calibrant solution, a reference solution and intermittent 
blood samples. Thus, during the continuous monitoring, for 

15 up to 36 hours, of a subjects blood gases, electrolytes 
and hematocrit levels, all measurements are performed with 
the sensors providing signal responses in the steady-state. 
No disclosure is included for deriving meaningful 
analytical information from solid-state electrodes before 

20 the electrodes attain an equilibrated "wet-up" state. 

U. S. Patent No, 4, 654, 127- issued to Baker and 
Funk discloses a sensing device equipped with species 
selective sensors and a rotatable multichamber reservoir in 
which calibrant and sample solutions are contained but in 

25 separate chambers. A plurality of chemical species may be 
detected by this device. However, as the sensors employed 
are not microf abricated, only a limited amount of control 
over the dimensions of the sensors ■ various components had 
been possible, resulting in their having nonuniform 

30 response behavior and necessitating the batch-wise 
determination of each sensor's response. The value of the 
response, e.g., intercept and slope, is then recorded on a 
bar code which must be read by a table top analyzer before 
the concentration of the desired chemical species may be 



WO 92/01928 



- 5 - 



PCT/US91/05071 



10 



calculated. Moreover, the disclosure of this patent is 
silent on methods for making useful measurements during 
wet-up of the sensing device. All indications, including 
those from available product literature, support the 
supposition that the- calibrant and sample solution 
measurements are carried out after the sensor response has 
attained a steady-state value. Furthermore, these 
commercially available sensors are stored in a high 
humidity package (i.e., substantially wet). This packaging 
method has the effect of limiting the shelf-life of these 
sensing devices significantly compared to competitive dry- 
reagent systems, such as those described below. However, 
this compromise ensures that the device is substantially 
"wet -up" at storage and, thus, enables the sensor to 
15 provide results fairly rapidly. Unfortunately, this 
compromise cuts back the device's useful shelf -life quite 
severely, particularly for enzyme-based sensing devices. 
Shelf-life, of course, can be extended to some extent by 
refrigerating the package. However, refrigeration adds to 
the expense of storage and also means that the device must 
be allowed to return to room temperature before use. 



20 



U. S. Patent Nos. 4,708,776 and 4,608,149 
disclose, on the other hand, improved "dry-operative" ion- 
selective electrodes. The inventors describe a " "dry- 

25 operative" electrode as "an ion-selective electrode which 
provides reproducible potent iometric determination [of] ion 
activity which is related to the ion concentration of an 
aqueous test solution with no requirement for wet storage 
or preconditioning prior to use" (col. 2, lines 10-15 of the 

30 second patent listed above). These patents also disclose 
methods of using such electrodes. In particular, the 
pbtentiometric determination of the concentration of sodium 
and potassium ions in an aqueous liquid is described. 
However, the method relies on a differential measurement 

35 which involves contacting the first of two "uniform" 



WO 92/01928 



- 6 - 



PCT/US91/05071 



electrodes i.e., a pair of identically formed potassium 
ion-selective electrodes, with the sample liquid and 
contacting the second electrode with a reference liquid 
(calibrant) containing a known amount of the ion of 
5 interest and then determining the the resulting difference 
between the two potential readings. This method requires 
that the sample and calibrant solutions be brought into 
contact with the respective electrodes strictly 
simultaneously to obtain reliable measurements of the 
10 analyte concentrations. Consequently, it is necessary to 
provide an automated means for the simultaneous application 
of the calibrant and sample fluids to prevent errors in the 
measurement . 

Pace, in European Patent Application No. 0 012 

15 035, describes self-calibrating miniaturized multiple 
sensors fabricated on a single chip. The usefulness of 
this disclosure is quite limited as the exact nature of the 
materials used for each of the multitude of layers 
described in the complex sensor structures is not revealed. 

20 Pairs of identical electrodes are used, a first member of 
the pair having at least two distinct electrolyte "layers" 
of known composition and the other, member of the pair 
either having no electrolyte present in its corresponding 
layers or having electrolyte present therein at a 

25 concentration which is significantly different from the 
first member. A discussion of the self-calibrating nature 
of these pairs of matched electrodes is present and makes 
clear that a differential method of signal measurement is 
employed to "nullify any drift and offsets in the 

30 measurement" (page 23, lines 30-31) m Moreover, this 
reference asserts further that these multiple layers which 
provide self-calibration "not only assure built-in 
reliability and accuracy, but relax manufacturing 
tolerances" (page 26, lines 1-4) . Thus, no successful 

35 means has been disclosed to manufacture simpler structures 



WO 92/01928 



- 7 - 



PCT/US91/05071 



with high dimensional tolerances nor has there been any 
suggestion that useful information may be derived from 
signal response measurements prior to attaining complete 
wet-up, much less that a method may be formulated in which 
5 such measurements are exploited. 

Accordingly, there remains at the present a need 
for a method which integrates a sensing device, preferably 
a microf abricated electrochemical sensor, having the 
requisite predictable, reproducible chemical response and 

10 "wet-up" characteristics, and an effective computational 
technique, which method allows the physician to obtain 
conveniently precise, accurate determinations of the 
concentration of analytes of clinical interest. Such 
determinations are desirably made in five minutes or less, 

15 most preferably within about a minute. 

2.2. PREVIOUS USES OF POTENTIAL PULSES 

Previous workers have utilized potential pulse 
techniques to increase the sensitivity of the 
elec chemical measurement or to reduce the flow 

20 dependence, of the electrode signal. However, these 
previous "applications have always involved fully wet-up 
devices utilized for the continuous monitoring of analyte 
concentrations. In such applications, previous workers 
sought to improve the signal output by taking their 

25 readings immediately after the application of a potential 
pulse across the sensor 'ectrodes. The cathode is^ open- 
circuited between* the pulses of applied potential . These 
and related techniques are described more fully in Short, 
D. L. and Shell, G. S. G. J. Phvs. E.?SrH ... Tnatnim. 19B5, 

30 13_ r 79-87 and Lilley, M. D. et al. J. Elertr oanal. Chpm. 
1969, 21, 425-429. 

On the other hand, methods exist for activating 
a catalytic surface including polishing away the surface 



WO 92/01928 



- 8 - 



PCI7US91/05071 



layer with a fine particle size, inert abrasive material 
such as alumina or placing the electrode in a corrosive 
acidic solution such as 1 molar sulfuric acid and cycling 
the applied potential for several minutes. Clearly, these 
5 existing methods are destructive and are inappropriate for 
the activation of an electrode surface overlaid with 
microf abricated biolayers . 

3 . SUMMARY OF THE INVENTION 

In accordance with the present invention a 
10 method is disclosed for determining the ratio (s) of the 
concentrations of preselected analyte species in more than 
one fluid which comprises, in part, providing 
microf abricated sensors having the requisite 
characteristics, which will be described more fully below, 
15 and performing signal measurements , before the equilibrated 
wet-up process is complete, while the sensors and reference 
electrode are in ; contact with a first fluid and, 
subsequently and separately, with a second fluid. 

In the present method, which fluid is brought 
20 into contaQt with the sensor first is unimportant, so long 
as a separate signal measurement is made while each fluid 
is in contact with the sensor and reference electrode. 
Despite the fact that, such measurements are taken before 
the sensors are fully w wet-up, n a process which may take 
25 several minutes, useful analytical, information about a 
variety of biological analytes can still, surprisingly, be 
obtained. The computational techniques for extracting this 
information from dry-stored microf abricated electrochemical 
sensors are disclosed. Thus, the present invention allows 
30 the simplicity and dry-storage capabilities of 
microfabricated sensors to be exploited while providing 
measurements of preselected analyte species as close to 
"real time" as possible. 



WO 92/01928 



- 9 - 



PCT/US91/05071 



It is thus an object of the present invention to 
provide a method of determining the concentration ratio of 
a preselected analyte ' species present in at least two 
fluids comprising (a) providing at least one 
5 microfabricated chemical sensor, which exhibits a response 
to changes in the concentration of a preselected analyte 
species, and a reference electrode capable of sustaining a 
well-behaved reference potential for a period of time 
sufficient to permit the completion of at least two signal 

10 measurements, against which reference potential of said 
sensor is measured, which sensor and reference electrode 
have been stored substantially dry, and which response is 
sufficiently rapid or "fast" relative to the "slow" 
monotonic wet-up behavior exhibited by said sensor and 

15 reference electrode when contacted with fluid; (b) 
establishing electrical contact between said sensor, 
reference electrode and external computational means; (c) 
contacting said sensor and reference electrode with a first 
fluid; (d) performing the first of said signal measurements 

20 in a preselected first time window in the presence of said 
first fluid; (e) displacing said first fluid; (f) 
contacting* said sensor and reference electrode with a 
second fluid; (g) performing the second of said signal 
measurements in a preselected second time window in the 

25 presence of said second fluid; and (h) relating said first 
and second signal measurements to the known concentration 
of said analyte species in one of said first or second 
fluids, to determine the unknown concentration of said 
analyte species in the other of said fluids before said 

30 sensor attains full equilibrated wet-up. 

It is also an object of the prese— invention to 
provide a method of determining the concentration of a 
plurality of preselected analyte species present in a 
sample fluid comprising, as a first step, providing an 
35 array of microfabricated potentiometric and amperometric 



WO 92/01928 



- 10 - 



PCT/US91/05071 



sensors, each sensitive to changes in the concentration of 
a particular preselected analyte species, and which array 
also comprises one or more reference electrode capable of 
sustaining a well-behaved reference potential for a 
5 sufficient period of time. Preferably, one reference 
electrode is dedicated to said potentiometric sensors and 
an other reference electrode is dedicated to said 
amperometric sensors. However, a single reference 
electrode may also be used for both types of sensors, if 

10 said amperometric sensors are also supplied with a common 
counter electrode. The counter electrode is designed to 
prevent polarization of the reference electrode, the effect 
of which polarization is more deleterious to the 
performance of potentiometric sensors. As noted, above, 

15 each sensor has been stored substantially dry and exhibits 
a response to said changes in the concentration of said 
particular preselected analyte species which is 
sufficiently rapid relative to the monotonic wet-up 
behavior of said sensors; (b) establishing electrical 

20 contact between said array of sensors and external 
computational means; (c) contacting said array of sensors 
with a first (e.g., calibrant) fluid; (d) performing a 
first set of signal measurements in 'a preselected first 
time window in the presence of said first fluid; (e) 

25 displacing said first fluid; (f ) contacting the array with 
a second (e.g., sample) fluid suspected of containing said 
plurality of analyte species, such that said array of 
sensors is in contact with said second fluid; (g) 
performing a second set of signal measurements in a 

30 preselected second time window in the presence of said 
second fluid; and (h) relating said first and second sets 
of signal response measurements to determine the 
concentration of a plurality of said preselected analyte 
species in said second fluid, based on the Lnown 

35 concentrations of each of said preselected analyte species 
in said first fluid. 



WO 92/01928 



- 11 - 



PCT/US91/05071 



It is a further object of the present invention 
to provide a method for activating the electrode surface 
of an amperometric sensor by subjecting the sensor to a 
series of potential changes while it is in contact with a 
5 fluid, preferably a calibrant fluid. 

Yet another object of the present invention 
involves providing a conductivity sensor by which the 
conductivity of the fluid in contact with the conductivity 
sensor may be determined and related, if desired, to the 
10 hematocrit level in the sample, or more simply to provide 
an indication of whether the fluid is calibrant, plasma, 
sr -urn or whole-blood, or even to provide a check on the 
pi sence of a sample of calibrant fluid. 

Still another object of the present invention 
15 includes a determination of the concentration of the 
analyte species of interest in about one mir e using dry- 
stored sensors . 

Yet another object of the present invention 
relates to minimizing the incidence of a test "failures" by 
20 incorporating data collection methods which allow the 
testing apparatus to scrutinize the integrity of the 
acquired signals and to manipulate the data set to exclude 
extraneous or aberrant data points which may otherwise lead 
to a rejection of a particular analysis. 

25 Other objects of the present invention should be 

readily apparent to those skilled in the art from the 
preceding discussion, as well as the following additional 
detailed disclosure. 

4 . BRIEF DESCRIPTION OF THE FIGURES 

W FIG. 1 illustrates the raw waveform exhibited by 

a microfabricated potentioroetric potassium ion sensor on 



WO 92/01928 



- 12 - 



PCT/US91/05071 



grounding (a) , exposure to a calibrant fluid (b) , initial 
sensor wet-up (c) , followed by transition to a sample fluid 
(e) . Suitable preselected time windows for data 
acquisition in each fluid are indicated by variable time 
5 segments (d) , for the first fluid, and (f ) , for the second 
fluid. 

FIG. 2a-2e show the response to a fluid change 
of a potassium ion sensor, sodium ion sensor, chloride ion 
sensor, urea sensor and microfabricated on-board reference 
10 electrode, respectively, with respect to an external 
standard Corning reference electrode. 

FIG. 3a-3d show the response to a fluid change 
of a potassium ion r sodium ion, chloride ion and urea 
sensors, respectively, with respect to a microfabricated 
15 on-board reference electrode. 

FIG. 4 §hows the signal output (current in 
namps) of an amperometric glucose sensor, suitable for use 
in the present method, in response to the applied electrode 
potential <mV) using a 20 mM glucose in HEPES buffer sample 
20 (O) or a HEPES buffer only (X) . 

FIG. 5a shows the increasing current output of a 
microfabricated glucose sensor as a series of activating 
pulse groups or potential changes are applied. 

FIG. 5b illustrates the response to a fluid 
25 change of the activated glucose sensor of FIG. 4a. 

FIG. 5c shows the absence of any substantial 
change in the current output of the activated glucose 
sensor of FIG. 4a upon application of additional pulse 
groups . 

30 FIG. 6 illustrates an electrochemical creatine 

kinase enzyme assay based on initial rate measurements. 



W^92/01928 



- 13 - 



PCT/US91/05071 



10 



15 



This figure also shows the sensor response to a fixed 
concentration of adenosine triphosphate (ATP), the product 
of the enzymatic process. 

5. DETAILED DESCRIPTION OF THE INVENTION 

The method of the present invention seeks to 
integrate the quantifiable, predictable behavior of 
microfabricated sensors during the equilibrated wet-up 
process with computational techniques, including data 
handling or collection methods, which may be implemented by 
the testing apparatus or external computational means, to 
arrive at a close estimate of the concentration of an 
analyte of interest in a given sample fluid. 

The "equilibrated wet-up" process is the means 
by which dry-stored sensors, by exposure to a fluid 
comprised of an aqueous medium, wet gas and the like, reach 
an operational state and, eventually, a steady-state. 
Here, the term "equilibrated wet-up" is used to enc.npass, 
not only ingress of water through the various membrane 
layers to the electrode surface, but also all of the 
physicochemical changes that occur prior to a sensor 
attaining that steady-state response. These changes and 
their consequences include: the hydration of each membrane 
layer and its effect upon migration therethrough of 
analytes, cof actors, ionopbores, enzyme, affinity-labels, 
25 and the like; the hydration of the enzyrae-cbntaining layer 
and its effect on the activity of enzymes or the 
selectivity of ionophores and affinity labels (i.e., the 
effect on their binding coefficients); the hydration of the 
electrode surface and the responsiveness of the sensor 
30 which may -be a function, for example, of the relative 
surface populations of metal oxide and metal hydroxide 
sites, or, alternatively, the degree of ligand-ligand 
substitution in which a silver-silver halide surface may be 
transformed into a halo-aquo metal complex. All of such 



20 



WO 92/01928 



- 14 - 



PCT/US91/05071 



changes that occur, before a fully wet-up, equilibrated 
sensor is obtained, can give rise either singularly or in 
concert, to a monotonically changing signal output despite 
the fact that the physical and chemical properties of the 
5 fluid, which may be comprised of an aqueous liquid or a wet 
gas, in contact with the sensor are not changing (i.e., the 
concentration of the preselected analyte in the fluid 
remains constant, along with the temperature, pH, osmotic 
pressure, ionic strength, etc., of the fluid, especially 
10 the calibrant fluid) . 

The processes described above may be considered 
passive in the sense that upon contact with fluid they 
occur spontaneously- However, the electrode surface of an 
amperometric sensor presents a special case in that its 

15 redox properties are substantially affected by the voltages 
that have been applied to it previously. In this sense, 
electrochemical activation of the electrode surface by 
applying a sequence of difference voltages can contribute 
significantly to the reduction of the time it takes before 

20 the sensor operates reliably in a steady-state manner, and 
therefore, contributes significantly to the apparent "wet- 
up" (or RC wet-up) of the sensor. 

The monotonic wet-up signal is usually described 
by a resistance capacitance (RC) time-constant because it 
25 can be simulated electronically by connecting, a resistor 
and capacitor either in series or in parallel. Thus, for a 
potential step applied to such a circuit, an exponentially 
decaying current, is obtained with a time-constant, T= RC 
where the current, i - E/Re - (t/t) . In this example the 

30 current necessary for changing the capacitor drops to 37 % 
of its original value at t ■ T and to 5 % of its original 
value at t * 3t. This latter value is often referred to as 
the 95 % response-time. For the sensors described here 
useful analytical information is obtained well before the 

35 sensors have transformed from the dry-state to a hydrated 



WO 92/01928 



- 15 - 



PCT/US91/05O71 



or wet-up state in which the sensors provide a signal which 
is 95 % of the expected equilibrated steady-state value. 

A good introductory discussion of this RC 
concept can be found in Lindner, E. et al. "Dynamic 
5 Characteristics of Ion-Selective Electrodes," CRC Press, 
1988 and the references cited therein, the complete 
disclosures of which are incorporated herein by reference. 
It should be noted that while the exponential model for 
time evaluation of the chemical sensor signal is usually 
10 referred to as ah "RC time-constant," no special detailed 
elucidation or assumptions about chemical or physical 
capacitive or resistive elements of the sensor's operation 
are required to verify the applicability of the exponential 
model. 

15 In the present invention, it has been 

surprisingly discovered that the RC time-constant for wet- 
up associated with the present wholly microf abricated 
sensors and reference electrodes can be manipulated and 
modeled closely. The reproducibility and predictability of 

20 this, wet-up RC time-constant is, in turn, a product of the 
microfabrication techniques described in the following 
section, and more fully in related co-pending U.S. 
Application Serial Nos. 07/432,714 and 07/245,102. Such 
microfabrication techniques provide much finer control over 

25 the dimensions of overlaid layers than can be achieved by 
previous techniques, including lamination. Indeed, a 
multiplicity of electrochemical sensors can now be 
fabricated on a single silicon wafer. These 
electrochemical sensors, including the first practical 

30 microfabricated reference electrode, possess sufficiently 
well-behaved properties to allow electrochemical 
measurements to be made with the precision and accuracy 
required in clinical chemistry. 



WO 92/01928 



- 16 - 



PCT/US91/05D71 



As will be described in more detail, below, the 
present inventors have discovered that the time evolution 
or rate of change in the monotonic wet-up signal can be 
predicted to a high degree of accuracy at any time after 
5 initial contact of the sensor and reference electrode with 
a fluid. This ability to model, analyze and manipulate the 
RC time-constant is an important aspect of the present 
method which allows the concentration of an analyte species 
of interest to be determined quickly and reliably before 
10 the sensors have attained a fully wet-up, equilibrated 
state. 

An equally important element of the present 
invention and one which is a direct consequence of having a 
reproducible, predictable wet-up RC constant is that the RC 
15 constant associated with the sensor's response to changes 
in the concentration of a preselected analyte is also 
highly predictable and precise. 

The RC time-constant, &F, is in units of seconds 
(sec) as evident from the following relationships: 

20 R (resistance) » £2, ohms » V/i 

C (capacitance in F, Faraday) - q/V - (i x sec)/V 

since, RC » flF, then 

RC = (V/i) x (i x sec)/V .» sec 

When a chemical sensor undergoes a change in its 
25 signal in response to a change in the analyte 
concentration, the exponential time constant governing the 
time dependence of the signal of a chemical sensor varies 
as taL 2 /D. fsee . Lindner, E. et al. y above, "Dynamic 

Characteristics of Ion-Selective Electrodes 11 , DRC Press, 
30 19B8 and earlier work by Buck, R.P. cited therein, 
especially Chapter 1 of "Ion-Selective Electrodes in 



WO 92/01928 



- 17 - 



PCT/US91/05071 



Analytical Chemistry", Freiser, H., (Ed), Plenum Press, 
1978). 

The time taken by a sensor to attain the steady- 
state response to a change in concentration of an analyte 
5 can be described by an RC time-constant. Thus, the 
RC r esponse time can be approximated by the relationship, RC = 
L 2 /D where, L is the thickness of the membrane layer in cm 
and D is the diffusion coefficient in cm 2 sec* 1 of the 
analyte through the membrane. The diffusion coefficient 

10 for glucose through a membrane layer is typically about 100 
times less than its diffusion coefficient in solution 
(Dsoln * 10" 6 cm 2 sec" 1 , D mem b * iO" 8 cm 2 sec* 1 ) , Thus, if 
the membrane layer is about 1 \im (10~ 4 cm) in thickness, 
then RC r esponse time is approximated to be (10- 4 ) 2 /10" 8 or 

15 ca. 1 sec. The RC W et-upr whose magnitude is on the order of 
tens of seconds, is thus "slow" relative to the response 
time. For example, the wet-up time-constants for the 
sensors shown in the Figures (e.g., FIGS. 1-3) exhibit X 

values of ca. 20 sec. As described below, when the wet-up 
20 process proceeds to an incomplete level where t<t<3t 

i.e., the rate of change in the signal is sufficiently 
small, it is possible to model this rate as a linear drift 
rate (S££, Table 1) . 

Clearly, the time required for a diffusion front 
25 of an analyte species to penetrate a membrane layer and 
establish a steady-state response increases with increasing 
membrane thickness. Thus, having a manufacturing method 
which attains a high degree of dimensional control over 
sensors 1 overlaid structures is crucial to obtaining 
30 devices with predictable, reproducible wet-up and response 
time behavior. 

In addition to describing a method involving a 
single sensor, the present invention is also concerned with 
the effective operation of an array of sensors each 



WO 92/01928 



- 18 - 



PCT/US91/05071 



10 



sensitive to a particular preselected analyte species. 
Each sensor in the array may be operating in one of a 
number of possible modes including, but not limited to, 
those designed to perform standard potentiometric, 
amperometric and conductimetric measurements or kinetic 
measurements based thereon. The concurrent operation of an 
array of different electrochemical sensors composed of 
some, or all, of these types, presents its own unique set 
of problems . 

Moreover, certain types of microelectrode 
assemblies are susceptible to inactivation, or a loss of 
surface catalytic activity, and require reactivation to 
secure the highest level of sensitivity. 

The following is a detail description of each 
15 element of the present analytical method for deriving the 
concentration of at least one, preferably a number, of 
preselected analyte 'species . 

5.1. WHOLLY MICROFABRICATED SENSORS 

Wholly microfabricated sensors, the availability 
20 of which comprises a preferred element of the present 
method, are described in detail in the applicants' prior 
co-pending U. S. Application Serial No. 07/432,714. 
Additional aspects related to the manufacture of integrated 
ambient sensing devices, including a microfabricated 
25 reference electrode, are described in US Patent No. 
4,739,380 and prior co-pending U. S. Application Serial No. 
07/156,262, the complete disclosures of which are 
incorporated herein by reference. These microfabricated 
sensors are manufactured in such a way as to avoid the 
30 errors and non-uniformity introduced by manual deposition 
of membranes and the like at various stages of the 
manufacturing process. Thus, a combination of. thin film 
techniques, including wafer-level photolithography and 



WO 92/01928 



- 19 - 



PCT/US91/05071 



automated microdispensing, are used to produce hundreds of 
identical sensors, or an array of different sensors, on a 
single silicon wafer. Reference electrodes made by the 
same process are also established in a highly controlled 
5 fashion. 

Such a reference electrode is necessary for 
electrochemical measurement of chemical or biochemical 
species in a sample solution ( See r Ives, D.V.G. and Janz, 
G.J. "Reference Electrodes, Theory and Practice, " Academic 

10 Press, 1961, the complete disclosure of which is 
incorporated herein by reference) . In the case of 
potent iometric measurement, the signal measured . the 
potential of a chemically responsive electrode (s^sor) 
with respect to the potential of the reference electrode. 

15 Ideally, the potential of the reference electrode is 
strictly independent of the chemical composition of the 
solution that it contacts. A reference electrode is also 
necessary for an amperometric measurement because it 
controls the potential of the amperometric sensor. Because 

20 such a high degree of control is present with regard to the 
composition of these layers, their physical dimensions, as 
well as their location on the .sensor array, the 
characteristics and specification of each sensor on the 
wafer, or any similarly produced wafer, are well-behaved 

25 and predictable. 

In particular, the microf abricated sensor and 
reference electrode which are most preferred comprises a 
permselective layer, superimposed over at least a portion 
of said sensor, having a thickness sufficient to exclude 

30 substantially molecules with a molecular weight of about 
120 or more while allowing the free permeation of molecules 
with a molecular weight of about 50 or less; and a biolayer 
superimposed over at least a portion of said permselective 
layvr and said sensor, ' which biolayer comprises (i) a 

35 sufficient amount of a bioactive molecule capable of 



WO 92/01928 PCT/US91/05071 

- 20 - 

selectively interacting with a particular analyte species, 
and (ii) a support matrix in which said bioactive molecule 
is incorporated, which matrix is derived from the group 
consisting of a photof ormable proteinaceous mixture, a 
5 film-forming latex, and combinations thereof and through 
which matrix said analyte species may freely permeate and 
interact with said bioactive molecule. 

In a preferred embodiment of the present 
invention the permselective layer is derived from a polymer 

10 film, most preferably comprising a heat-treated film of a 
silane compound having the formula R'nSi (OR) 4-m in which n 
is an integer selected from the group consisting of 0, 1, 
and 2; R f is a hydrocarbon radical comprising 3-12 carbon 
atoms; and R is a hydrogen radical or a lower alkyl radical 

15 comprising 1-4 carbon atoms. 

The bioactive molecule of the sensor biolayer 
may be selected from a wide variety of molecules well known 
to those skilled in the art and may include, for example, 
an ionophore, an enzyme, a protein, polypeptide, nucleic 
20 acid or an immunoreactive molecule. Typically, the 
bioactive molecule is an ionophore or an enzyme. 

In another preferred embodiment of the 
microfabricated sensor, the photof ormable proteinaceous 
mixture comprises (i) a proteinaceous substance; (ii) an 

25 effective amount of a photosensitizer uniformly dispersed 
in said proteinaceous substance; and (iii) water. Examples 
of proteinaceous substances which are useful in the present 
invention include albumin, casein, gamma-globulin, 
collagen/ derivatives, and mixtures thereof. The most 

30 preferred proteinaceous substance is an animal gelatin, 
especially fish gelatin. Many types of photosensitizers 
abound. Of particular interest, however, are high 
oxidation state transition metal compounds, especially iron 
and chromium salts. 



WO 92/01928 



PCI7US91/05071 



- 21 - 



The film-forming latex may comprise an aqueous 
emulsion of a polymer or copolymer derived from synthetic 
or natural sources. 

Of course, additional layers may be present in 
5 the microfabricated sensors. For example, additional 
layers may be used to attenuate the transport of selected 
molecules, including analyte species, through the sensor. 
An electrolyte layer may be present especially for the 
potentiometric sensors or the reference electrode 
10 structure. A more complete description of the reference 
electrode structure may be found in applicants' prior co- 
pending U. S. Application Nos. 07/432,714 and 07/156,262. 

It should be re-emphasized that an array of 
sensors may use a common reference electrode. Thus, a 

15 serieis of potentiometric sensors may be assembled for 
measuring the activity of several electrolytes 
concurrently, the signal of each sensor being determined 
relative to the potential of the common reference 
electrode. Amperometric sensors may have a slightly 

20 different configuration, each comprising a sensor and a 
counter electrode, for example, but with all the sensors in 
an array sharing a common reference electrode. 

The conductivity sensor for a hematocrit 
measurement is plain in design comprising two noble metal 

25 electrodes spaced at an appropriate distance on the 
proposed array. In carrying out a conductivity 
measurement, an electric field is generated between the 
pair of metal electrodes by applying an a.c. signal (a d.c, 
signal may also be employed). Preferably, the effect of 

30 the field is limited predominantly to the fluid compartment 
directly above the pair of metal electrodes. This 
configuration maximizes the device's sensitivity toward 
erythrocytes. It should be noted that several factors need 



WO 92/01928 



- 22 - 



PCT/US91/05071 



to be taken into account in selecting the appropriate 
frequency for the a.c. signal. These factors include 
minimizing Faradaic processes at the electrode surface 
while maximizing the distortion of the field by the 
5 erythrocytes . 

The electrochemical sensors which are perhaps 
the most complex, in terms of the need for additional 
reagents, are those sensors used to perform kinetic 
measurements. These types of sensors are useful in 

10 determining, for instance, the activity of an enzyme as 
reflected by the rate of change in the concentration of a 
detectable species consumed or produced by the enzyme- 
linked reaction. Hence, the activity of a particular 
enzyme in a given sample may be established. Also, certain 

15 enzyme-linked immunoassays may be carried out, paving the 
way for the analysis of a wide variety of immunoreactive 
and affinity-active species, including antigens, haptens, 
antibodies, viruses and the like. 

Thus, a preferred embodiment of a sensor 
20 intended for enzyme or immunoassays should have a layer, 
accessible to the sample fluid, to which is immobilized one 
or the other of a ligand/ligand receptor pair. Again, the 
reader is referred to the disclosure of applicants 1 prior 
co-pending U. S. Application Serial No. 07/432,714 for 
25 further details. 

5.1.1. DISPOSABLE DEVICE FOR SENSORS 

The microfabricated sensors described above are 
preferably contained in a disposable device which can be 
adapted for performing a variety of measurements on blood 
30 or other fluids . The disposable device is constructed to 
serve a multiplicity of functions including sample 
collection and retention, sensor calibration and 
measurement. During operation, the disposable device may 



WO 92/01928 



- 23 - 



PCT/US91/05071 



be inserted into a hand-held reader which provides the 
electrical connections to the sensors and automatically 
controls the measurement sequence without operator 
intervention, 

5 A suitable disposable device includes upper and 

lower housing members in which are mounted a plurality of 
sensors and electrical contacts and a pouch containing a 
calibrant fluid. The sensors generate electric potentials 
based on the concentration of specific ionic species in the 
10 fluid sample tested. A double sided adhesive sheet is 
situated between the upper and lower housing members to 
bond the housing members together and to define and seal 
several cavities and conduits in the device. 

A first cavity is located at the center of the 
15 device having a pin at the bottom of the cavity and a 
hinged disc at the top of cavity. A sealed pouch 
containing calibrant fluid resides in the cavity and a 
first conduit leads from this cavity toward the sensors. 
A second conduit has orifice at one end for the receipt of 
20 a fluid sample while the other end of the tube terminates 
at a capillary break. A third conduit leads from the 
capillary break across the sensors * to a second cavity 
which serves as a sink. The first conduit joins the third 
conduit after the capillary break and before the sensors. 
25 A third cavity functions as an air bladder. When the air 
bladder is depressed, the air is forced down a fourth 
conduit into the second conduit. 

In operation, a fluid sample is drawn into the 
second conduit by capillary action by putting the orifice 
30 at one end of the conduit in contact with the sample. 
After the sample fills the second conduit, the orifice is 
sealed off. The pouch containing the calibrant fluid is 
then pierced by depressing the disc down on the pouch which 
causes the pin to pierce the other side of the pouch. Once 



WO 92/01928 



- 24 - 



PCT/US91/05071 



the pouch is pierced, the calibrant fluid flows from the 
cavity through the first conduit to the third conduit and 
across the sensors at which time the sensor calibration is 
performed. Next, the air bladder is depressed forcing air 
5 down the fourth conduit to one end of the second conduit 
which forces the sample out the other end of the conduit, 
past the capillary break, and into the third conduit and 
across the sensors where measurements are performed. As 
this is done, the calibration fluid is forced out the third 
10 conduit into the second cavity where it is held. Once the 
measurements are made r the disposable device, can be 
discarded. 

The hand-held reader includes an opening in 
which the disposable device is received, and a series of 

15 ramps which control the test sequence and the flow of the 
fluid across the sensors. As the disposable device is 
inserted into the reader, the reader ruptures the pouch of 
calibrant fluid by depressing the hinged disc. The reader 
then engages the electrical contacts on the disposable 

20 device, calibrates the sensors, depresses the air bladder 
to force the fluid sample across the sensors, records and 
electric potentials produced by the sensors, calculates the 
concentration of the chemical species tested and displays 
the information for use in medical evaluation and 

25 diagnosis. 

Thus, for example, to measure the potassium 
concentration of a patient's blood, the physician or 
technician pricks the . patient f s finger to draw a small 
amount of blood. The physician then puts the orifice of 

30 the device into the blood, drawing the blood into the 
device through capillary action. The physician then seals 
off the orifice and inserts the device into the reader. 
Upon insertion, a sequence of events is automatically 
initiated by the reader without intervention from the 

35 physician. The reader automatically causes the calibrant 



WO 92/01928 



PCT/US91/05071 



- 25 - 

pouch to be punctured so that the calibrant fluid flows 
over the sensors, activating the sensors and providing the 
necessary fluid for calibration. The electrical contacts 
of the device are then automatically connected to the 
5 reader and the calibration measurements are automatically 
made. The reader then automatically depresses the air 
bladder in the disposable device causing the sample to flow 
over the sensors. The electric potentials generated by the 
sensors are read and the concentration of the chemical 
10 species is automatically calculated. The result is 
displayed or output to a printer for the physician to 
utilize. 

Upon completion of the process, the physician 
removes the device from the reader and disposes of it 
15 properly. The reader is then ready to perform another 
measurement which is initiated by the insertion of another 
disposable device. 

5.2. DATA HANDLING METHODS FOR PERFORMING 
SIGNAL RESPONSE MEASUREMENTS 

20 The present data handling methods allow the 

instrument housing the external computational, data storage 
and display means to extract the needed information (i.e., 
the electrochemical response of each sensor in the array) 
from a background which includes sensor wet-up, fluidics 

25 transients (those transients associated with fluid flow) 
electronic noise, contact noise associated with the 
electronic interface between the sensor and reference 
electrode outputs and the computation means and other 
intermittent artifacts or signal fluctuations. The reader 

30 is referred to applicants 1 prior co-pending U. S. 
Application Serial No. 07/187,665 for further information 
concerning salient components of a handrheld instrument 
which may be used for processing the sensors signals. 



WO 92/01928 



- 26 - 



PCT/US91/05071 



The data handling techniques include 
computational methods which are designed to relate the 
first and second signal measurement and, thus f provides the 
concentration ratios of the preselected analyte species in 
5 the first and second fluids. These computational methods 
distinguish the relatively fast response of the sensor 
(either potentiometric or amperometric) to concentration 
changes from the slower monotonic wet-up of both the 
sensors and reference electrode. After studying the 

10 particular embodiments described herein, it should be 
apparent to one skilled in the art, however, that more 
sophisticated computational methods may be employed to 
process signals, should non-monotonic wet-up behavior be 
. encountered. Still other types of computational methods 

15 may hereafter be conceived for detecting signal defects as 
discussed below. In any event, a first order approximation 
can be extended to a more general nth order polynomial 
relationship, exponential relationship and the like, if the 
need is apparent. The only concern with higher order 

20 computations is their suitability for extrapolation. 

Hence, in one embodiment of the present method a 
computational method is employed to detect unusable signals 
caused by: changes in the nature of the fluid in contact 
with the sensor, electrical noise from the contacts or 

25 connections between the individual sensors and the external 
computational means, as well as other extraneous 
intermittent artifacts. The present computational methods 
may be used to provide an indication of the occurrence of 
artifacts which combine to give an unacceptable measurement 

30 cycle. In a preferred embodiment of the present method, 
the standard computational method is extended to allow the 
instrument to remove the offending artifacts or aberrant 
data point from the acquired data set. The "corrected" 
subset is then processed in the same way to provide a 

35 useful measurement. Thus, certain but not all analysis, 



WO 92/01928 



- 27 - 



PCT/US91/05071 



which otherwise would have been di,. warded as "failures," 
are salvaged under appropriate conditions. 

It should be pointed out that the computational 
methods employed in the present method, though similar to 
5 known signal processing methods, perform, inter alia, a 
non-trivial assessment of how the electrochemical response 
of interest and the wet-up behavior or artifacts contribute 
to the raw waveforms. The desired signals must then be 
derived or further manipulated before useful information is 
10 obtained. 

The present data handling method can be broken 
down into two main parts comprising a data acquisition 
portion and another for data manipulation and analysis. 
Each portion has. its own set of computational methods, and 
15 their relationship may be better understood by referring to 
Chart I, below. 



WO 92/01928 



- 28 - 



PCT/US91/0507! 



CHART I 



DATA ACQUISITION 



1 SIGNAL CURRENTS 1 | SIGNAL VOLTAGE S? 



ACTIVATION ALGORITHM 
AND ELECTRONICS 



ELECTRONICS 
AND SOFTWARE 



DATA MANIPULATION &HB ANALYSIS 







| DATA MEMORY 







UNACCEPTABLE DATA 
NOT DISPLAYED 



CORRUPT DATA 
RECOVERY 




SIGNATURE 
ANALYSIS 



ACCEPTABLE DATA 
CALCULATION 



DISPLAY ANSWER 



WO 92/01928 



- 29 - 



PCI7US91/05071 



10 



During the data acquisition portion of the data 
handling method, the analog signals obtained from the 
sensors are converted into a digital format for recording 
in the data memory. The electronics are designed to be 
appropriate for high impedance potentiometric sensors with 
sufficient resolution over the expected range of voltage 
measurements. The electronics for the amperometric 
sensors, which measure current, include current to voltage 
converters and are also designed to have sufficient 
resolution over the expected range of current measurements. 
During the measurements, the fluid is grounded so as to 
prevent the fluid potential from floating out of the range 
of the operational amplifiers. 

It should be pointed out that the amperometric 
15 sensors are preferably subjected to an electrochemical 
activation process . Applicants speculate that this 
activation process enhances the catalytic activity of the 
sensor electrode surface toward reduction or oxidation of 
certain redox active chemical species. Further discussion 
on this electrical activation process is presented in a 
later section, below. 



20 



Also, this portion of the data handling method 
includes the collection and digital storage of conductivity 
measurements. As mentioned elsewhere in this disclosure, 
25 these conductivity measurements are related to the analysis 
of the patient's hematocrit levels and, also, system 
quality assurance methods. It is important to note that 
the wet-up of conductimetric sensors as described herein is 
extremely fast (on the order of milliseconds) because the 
sensor is comprised simply of two metal electrodes directly 
in contact with the fluid. Hence, no extended wet-up of 
intervening membranes is observed. 



30 



WO 92/01928 



- 30 - 



PCI7US91/05071 



Additional aspects of the data acquisition, 
manipulation and analysis are discussed in further detail 
in a later section, below. 

5.2.1. SIGNATURE ANALYSIS 

5 The signature analysis section of Chart I is the 

step in which the integrity of the signal measurements, 
performed in the first and second data time windows, is 
analyzed. The computational methods used in this section 
detect the presence of excessive wet-up RC time constants, 
10 spikes, glitches and noise and compare the values observed 
within the time windows with preferred values held in 
memory . 

In a specific embodiment of the present method, 
the first part of the signature analysis is run in real 

15 time (i.e., during data collection) in which a seven point 
sliding window slope analysis is implemented, beginning 
with the first seven points of the time window. A total of 
twenty-five data points (1-25) are actually collected in 
each time window, whether the measurements are taken in the 

20 presence of the first or second fluids. Although any size 
can be chosen for the sliding windows, the seven point 
sliding window provides an acceptable level of resolution. 

The computational method is applied to each 
sensor during data collection at each time window. In an 

25 embodiment which utilizes an array of sensors, the 
computational method keeps track of which sensor is active 
and whether or not the time window has begun. The slope of 
each seven point window is computed, based on a recursive 
form of a linear regression. As each new point is 

30 collected, causing the seven point window to slide, the 
slope of the new seven point window (points 2-8) is 
compared with the first seven point window (points 1-7) . A 
range of acceptable values for the new slope based on the 



WO 92/01928 



- 31 - 



PCT/US91/05071 



value of the first slope (the basis slope) can be set, and 
if any window is found to fall outside that range, a bit is 
set for that sensor as a flag for later use. • 

The present sliding window slope analysis 
5 detects spikes and glitches in the time windows by looking 
at first derivative changes. It should be apparent that 
the slope of each new window can be compared with the slope 
of the window which immediately preceded it instead of 
comparing the new slope with that of the first seven point 

10 window. The latter option saves time, however, by avoiding 
the need to save in memory a new value for the basis slope 
with each pass through the computational method. Also, the 
present technique is more sensitive to low frequency 
glitches than the alternate approach which involves a 

15 trade-off in determining signal integrity. Yet another 
alternative method could compare the slope of both nearest 
neighbors rather than just the preceding point. Other 
methods should be apparent to those skilled in the art. 

The second part of the signature analysis 

20 section involves a post-data collection processing 
computational method that checks for limits on the observed 
values. These computational methods may include the 
calculation of first derivatives, error or estimates of 
linear fit, delta drift rates, mean drift rates, second 

25 derivatives, degree of curvature and the like. That is, 
the computational method compares the observed data with 
the expected range of values held in memory. In a specific 
embodiment of the present method, limits are placed on the 
drift rate (the slope) of each time window, the difference 

30 between each time window's drift rate (i.e., the delta 
drift rate between the first fluid time window and the 
second fluid time window) and the mean or average value of 
the response obtained in each time window. The drift rate 
and mean values ar£ obtained from a linear regression 

35 analysis. Maximum and minimum set values for each 



WO 92/01928 PCT/US91/05071 

- 32 - 



parameter may be different for each sensor. Moreover, an 
additional section of the post-data collection signature 
analysis computational method computes the error associated 
with the least squares fit and compares the value of this 
error with the limit placed for each sensor in the array. 
It is the selection of an optional combination of limits 
that determines the accuracy and precision that is 
attainable. For example tightening the noise limits but 
relaxing the wet-up RC time constant may be preferable over 
the reverse procedure. 

Subsequent sections of the data manipulation and 
analysis portion of the data handling package dictates 
further actions if a particular sensors' time window(s) 
contains spikes, glitches, noise or observed values (e.g., 
15 drift rates, delta drift rates, mean values or error of 
estimates) falling outside an expected range. The results 
of the affected sensors are not displayed and appropriate 
advisories are then displayed over the instrument monitor. 

However, if the data contain manageable 

20 aberrations, a corrupt data recovery computational method 
is then employed to derive sufficient information to 
deliver a useful result. The corrupt data recovery 
computational method includes a determination of whether 
the detected glitch and/or spike is sufficiently large to 

25 affect deleteriously the propriety of the linear fit 
applied to the entire time window. This analysis is 
accomplished by comparing the linear fit applied to the 
entire time window to the basis slope obtained from the 
first sliding seven point window. Alternatively, the 

30 linear fits, with and without the offending glitch removed, 
imay be compared and a decision made regarding its overall 
integrity. Still another recovery computational method 
implements a type of median filter to smooth detected 
glitches. Again, other recovery methods may be 

35 contemplated . 



WO 92/01928 



- 33 - 



PCT/US91/05071 



It may be useful, at this stage in the 
discussion, to describe briefly the behavior of a 
representative microfabricated sensor in the course of a 
typical fluid analysis cycle. Figure 1 illustrates the 
5 potential response, as a function of time, of a 
potent iometric potassium ion sensor utilized in the present 
invention. During the first several seconds, the 
measurement cycle is initiated electronically, contact 
being made between the array of sensors, which are mounted 

10 preferably on a disposable assembly for example, 

prior co-pending U.S. Application Serial No. 07/245,102, 
the complete disclosure of which is incorporated herein by 
reference), and the external computational means. During 
this initial period, all the sensors are grounded. Within 

15 a few seconds after the initial electrical contact is 
established, the first fluid is caused to flow over and 
make contact with the sensor array. As wet-up ensues the 
potential of the potent iometric sensor drifts 
monotonically, essentially in an exponential manner. 

20 In the particular sequence illustrated in Figure 

1, the fluid change is made after about 72 seconds, though 
clearly, the introduction of the second fluid can be made 
much sooner if obtaining this potassium ion concentration 
is the only objective of the analytical cycle. In the 

25 present illustration, the first time window can be selected 
to fall at any suitable time after the first fluid is in 
place over the sensors and prior to the introduction of the 
second fluid. The second time window, in which the second 
set of signal measurements is performed, is begun 

30 preferably soon after the fluid change is made. 



WO 92/01928 



- 34 - 



PCT/US91/05071 



5.2.2, CALCULATION OF THE 
SENSOR RESPONSE 

Calculation of the sensor response, that is, the 
change in potential reading between the first and second 
5 fluid (the delta voltage) , is performed once an adequate 
data set has been obtained. The present algorithm used to 
calculate the sensor's response performs two linear least- 
squares fits . One least-squares fit is performed in the 
first fluid time window; the other is done in the second 
10 fluid time window. 

There exists many different ways to extract the 
sensor response from a raw waveform such as that shown in 
Figure 1. Some examples for deriving the delta response 
value include, but are not limited to, a linear/linear 

15 extrapolation, a linear/mean calculation, a mean/mean 
approach or a mean/linear method, to name a few. In the 
linear/linear case, the fit to the first time window is 
extrapolated forward to an estimated fluid transition 
point, the middle of the transition period. The fit to the 

20 second time window is extrapolated backward to the same 
estimated transition point, although clearly, it is not 
necessary to specifically select the fluid transition point 
at the midpoint . as some other suitable point at the may 
also be selected) . The arithmetic difference between these 

25 extrapolated voltages is the sensor's delta response. 
Alternatively, the ratio between the extrapolated values is 
calculated if currents derived from an amperometric sensor 
are being measured. The concentration in the sample fluid 
of the analyte species of interest can then be determined 

30 based on the known chemical activity (concentration) of the 
analyte in the calibrant fluid. The Nicolsky extension (1) 
of the Nernst equation, which also takes into account the 
effect on the sensor response of interfering ions present 



WO 92/01928 



- 35 - 



PCT/US91/05071 



in the fluids, is used for this determination, in the case 
of potentiometric sensors: 

E - E 0 + RT/nF log [A + k a ,h B] (1) 

where E is the measured electromotive force (signal) , R is 
5 the gas law constant, T is the absolute temperature, n is 
the absolute value of the charge on analyte species a_ 
(e.g., n = 1 for the ammonium ion), F is the Faraday 
constant, A is the activity of the analyte species a, B is 
the activity of an interfering chemical species h, k a . h is 

10 the interference coefficient associated with the effect of 
the presence of chemical species h on the electrochemical 
potentiometric determination of the activity of the analyte 
species and E D is a constant independent of T, A or B. 
For additional discussion of the Nicolsky equation, please 

15 refer to Amman, D. Ion-Selective Mir roelectrodfis r Springer, 
Berlin (1986) p. 68 and references cited therein, the 
complete disclosures of which are incorporated herein by 
reference. 

A correction of the calculated chemical activity 
20 of the analyte in the unknown or sample fluid can be 
obtained by applying the Henderson equation (2) where the 
term Ej is included in equation (1) and taking into 
consideration differences in ionic strength and matrix 
effects between the calibrant and sample fluids, which 
25 differences are usually manifested as a slight response of 
the reference electrode: 

Xi 1 1 Zj I Hi/zi3 [ci (p) -c i (tt) ) RT ZilZiljiiCi ta) 

Ej = — ; In — (2) 

liUzilHi] [ci(p)-Ci(a)] F liUimiCifp) 

where zi is the charge, \i± is the mobility, Ci is the molar 
concentration of species i and a and p are transfer 
30 coefficients. 



WO 92/01928 



- 36 - 



PCT/US91/05071 



For amperometric sensors , the concentration of 
the analyte in the sample fluid is derived from the 
measured current ratio r the known concentration of the 
preselected analyte in the calibrant fluid and a system 
5 constant. Equation 3 may thus be used: 

[Ux/i2)ciY] + 5 « c 2 (3) 

where ii and ±2 are the measured currents of the first and 
second fluids and ci and C2 are molar concentrations of a 
preselected analyte. If the first fluid is calibrant, then 
10 ci is known and the value of C2 in the sample can be 
readily obtained. The correction factors y and 8 are 
derived experimentally and take into account differences in 
the physicochemical properties of the calibrant and sample 
fluids. 

!5 m the linear/mean approach, the linear fit to 

the calibrant time window is extrapolated forward, as 
before, but to the midpoint of the sample time window. The 
difference between the extrapolated value from the 
calibrant fit and the mean sample value is the delta 

20 response. Again, the ratio of the extrapolated value of 
one time window to the midpoint value of the second time 
window can be calculated also, if desired. The mean/linear 
case reverses the direction of the extrapolation used in 
the preceding approach, and the execution of the mean/mean 

25 method of calculation should be fairly evident. 

Ultimately, the method of choice for calculating 
the sensor response depends on the characteristics of that 
sensor and may be best determined through routine 
experimentation. However, the quality of the sensor 
30 response measurements will certainly have an impact on 
which method is most appropriate. For instance, a frirly 
large difference between the slopes of the data points in 
the first and second time windows may indicate that a 



WO 92/01928 



- 37 - 



PCT/US91/0507! 



linear method may lead to a skewed result and that a mean 
value of the particular time window may be more 
appropriate. As already discussed, above, the 
computational methods utilized during data collection, 
5 manipulation and analysis may be enhanced to look at drift 
rates (slopes) as data points are collected in each time 
window to determine the "smoothness" of the window. 
Likewise, first and second derivatives of fully sampled 
time windows can be compared using "enhanced" computational 
10 methods. 

5.3. C A LI BRANT FLUID 

The calibrant fluid should contain a known 
concentration of the preselected analyte. The calibrant 
fluid may comprise a wet gas but is preferably comprised of 

15 an aqueous liquid. The chosen concentration of the 
preselected analyte in the calibrant is preferably similar 
to that expected to be encountered in the unknown sample. 
When necessary, preservatives (e.g., p-hydroxybenzoate, 
phenylmercuricacetate, p-aminobenzaldehyde and the like) 

20 may also be included to prevent microbial contamination. 
In the preferred method of the present invention, the 
calibrant fluid is an aqueous solution of several analytes 
whose concentrations are similar to those expected to be 
determined in an unknown sample, usually whole blood/ It 

25 has been the experience of the inventors that, due to 
differences in the properties (e.g., viscosity) of the two 
fluids, the practice of the present method is simplified if 
a small volume* of blood- is used to displace a small volume 
of calibrant fluid. That is, the first fluid is preferably 

30 the calibrant,- and the second fluid is preferably the blood 
sample. It should be evident, however, that the present 
method is not so limited,, and the sequence and nature of 
the fluids introduced to the sensors is a matter of choice. 



WO 92/01928 



- 38 - 



PCT/US91/05071 



5.4. POTENTIOMETRIC SIGNAL RESPONSE 

Although the signal response phenomena of an 
ion-selective electrode (ISE) upon change of the chemical 
composition of the fluid with which the electrode makes 
5 contact have been the object of much investigation in the 
chemical literature, the time-dependent signal of a dry- 
stored ISE upon first contact (wet-up) with an aqueous 
medium has received relatively little attention. As stated 
previously, this neglect results from the generally 
10 accepted view in the art that prior to attaining a complete 
equilibrated wet-up, the ISE can be of no analytical value. 
However, while this premise may hold true for 
conventionally fabricated macroelectrodes, it is not 
necessarily true for microfabricated devices. 

15 Clearly, to obtain useful measurements as close 

to "real time" as possible, it would be very desirable to 
record analytical readings from dry-stored ISEs before the 
equilibrated wet-up process is complete. One means for 
obtaining this measurement involves, the previously 

20 described prior art differential method in which a pair of 
electrodes with the same structure is used for each analyte 
species and corresponding calibrant solution . . In such a 
configuration, there is no conventional reference 
electrode . The differential reading yields the ratio of 

25 unknowns o-known concentrations with regard to that 
particular chemical species to which that ISE is primarily 
responsive. Based on the assumption that each of the pair 
of " ISEs is identical in the characteristics that give rise 
to the wet-up RC time-constant, e.g., physical dimension 

30 and material composition, the expectation that the 
monotonic wet-up signal will be canceled out of a 
differential measurement appears to be reasonable. 
Although an approach of some utility, it has certain 
inherent limitations with regard to rapid, multi-species 

35 electrochemical assays in parallel because it requires 



WO 92/01928 



- 39 - 



PCT/US91/05071 



Strictly simultaneous contact between each ISE of each 
matched pair and the calibrant and sample fluids, 
respectively. The problems associated with conducting and 
using an array of these matched pairs of ISEs for the 
5 analysis of multiple species would appear prohibitive as 
these pairs have been used only for single analyte 
determinations. 

By the present method/ such limitations are 
eliminated. The wet-up phenomena can be controlled to a 

10 high degree of confidence, along with the physical 
dimensions and material properties of the sensors and 
reference electrode. In the present invention the output 
shows that the wet-up has proceeded to a level where x < t 
< 3x the slow rate of change can be treated as a linear 

15 drift and this linear relationship can reliably project 
the sensor's output forward in the time domain for some 
brief period, e.g.,. 30 seconds. This capability becomes 
crucial when one changes the solution in contact with the 
sensor and reference electrode, e.g., (unknown) sample 

20 (known) calibrant and compare the readings between the two 
fluids. In this manner the need for strictly simultaneous 
introduction of calibrant to the "reference" electrode and 
the unknown sample to the "working" electrode is 
circumvented. 

25 It should be pointed out that even though a 

mlcrof abdicated reference electrode is likely to have its 
reference quality compromised by the eventual inward or 
outward diffusion of ionic species from the aqueous medium, 
this process is typically monotonic in the longer time 

30 domain. As a consequence, output of a sensor measured 
against such a reference electrode will be modeled reliably 
by a linear relationship. Linear relationships are 
generally preferred over higher order polynomials because 
of the basic simplicity, less vulnerability of the signal 

35 to interruptions, and greater confidence in obtaining a 



WO 92/01928 



- 40 - 



PCT/US91/05071 



meaningful extrapolation involved with the former. 
Furthermore, it should be noted that the general 
exponential time dependence of such relaxation phenomena 
can be shown mathematically to be well approximated by a 
5 linear function at sufficiently large values of time . The 
propriety of this approximation becomes apparent if one 
notes that the Taylor series expansion of the experimental 
function is dominated by the linear terms as the argument 
(-t/T) approaches (-a). 

A series of Figures (1-3) follows to illustrate 
the wet-up dynamics of several different wholly 
microfabricated sensors with respect to a fully wet-up 
conventional flow through silver-silver chloride reference 
electrode, as well as the performance of such sensors 
versus a microfabricated reference electrode that undergoes 
similar wet -up effects. It can be seen that a linear fit 
of the reading in one fluid enables one to predict 
accurately what the value of the reading should be at later 
times for comparison with a subsequent reading in a second 
fluid. 

•It is important to characterize the wet-up 
behavior of a dry-stored, thin-film ion selective electrode 
upon exposure to a calibrant fluid. As illustrated in 
Figure 1, after the grounding path is eliminated and a 
25 fluid path is established between the potassium sensor and 
the reference electrode (at about 9 seconds interval) , the 
raw waveform becomes more manageable. The exponential 
decay of the potential output toward a steady-state value 
is. a function of the wet-up of the potassium sensor and its 
30 inherent RC time constant. 

Figures 2a to 2e show the response 
characteristics for a potassium, sodium, chloride, urea and 
on-board microfabricated reference electrode, respectively, 
versus a standard Corning reference electrode, with the 



WO 92/01928 



- 41 - 



PCT/US91/05071 



change from calibrant to the unknown (sample) fluid 
occurring after about 50 seconds. (In these and subsequent 
Figures the response of the sensor prior to introduction of 
the calibrant fluid is not shown.) The wet-up response for 
5 all of these ISEs is similar, showing a roughly linear 
drift rate after about 40 seconds. It is particularly 
important to note that the on-board microf abricated 
reference electrode (Figure 2e) does not respond to the 
fluid change, beyond minor correctable ionic strength and 

10 matrix effects. Hence it may be used advantageously as the 
actual reference electrode for the potassium, sodium, 
chloride, and urea sensors. In Figures 3a-3d, the data 
corresponding to these respective sensors in which the 
microfabricated reference electrode is used as the on-board 

15 reference demonstrates that, indeed, the on-board reference 
electrode operates perfunctorily and that the glitches 
associated with fluid change in Figures 2a-2d are even 
eliminated. This result is observed due, presumably, to 
closer proximity of the on-board reference electrodes to 

20 the sensors. 

In terms of the selection criteria for the type 
of computational method for data manipulation and analysis, 
an empirical approach can be employed which involves 
determining the accuracy and precision for each of the 

25 first order relationships. A Linear/Linear method is found 
to be superior for the potassium, sodium and chloride 
sensors. The urea sensor, which has a slower response- 
time, provides best results when a Linear /Mean fit is used. 
The exact location and duration of the data acquisition 

30 windows can also be determined in this fashion, 

5.5. AMPEROMETRIC SIGNAL RESPONSE 

Unlike potentiometric measurements where the 
ideal sensitivity of the response of a sensor is determined 
by fundamental constants and intensive thermodynamic 



WO 92/01928 



42 - 



PCT/US91/05071 



quantities via the Nernst equation, the current output in 
amperometric measurements is system dependent. That is, 
the result is dependent upon the geometry and transport 
properties of the overlaid structures, as well as the 
5 surface properties of the electrode . However, over some 
specific concentration range the current response will be 
directly proportional to the bulk concentration of the 
analyte. However the absolute current at any given bulk 
concentration usually increases over the time domain, as 

10 the dry-stored overlaid structures hydrate; that is, the 
transport rate across the membranes increases. This 
process can compromise the analytical value of the sensor, 
limiting its achievable accuracy and precision, unless the 
calibration process and subsequent measurement in a sample 

15 (unknown) fluid are performed close together over the time 
domain. Clearly, the operation of dry-stored, single-use 
amperometric sensors require that the sequence of steps 
performed over the course of the measurement be controlled 
in a careful fashion. 

20 In a preferred embodiment of the present 

invention, at least two amperometric sensor signals (e.g., 
current output) are determined in each data aquisition time 
window (again, there should be at least two time windows, 
one for each fluid) . Most preferably, one of the sensor 

25 signals determined in each time window (e.g., the calibrant 
fluid time window) is measured at a first applied potential 
and the other of the sensor signals, still in the same time 
window, is measured at a second applied potential. A fluid 
change is then made, and the above signal measuring process 

30 is repeated.: If, for instance, the applied potential is 
stepped up from one value to a higher applied potential in 
the first time window, then in the second time window, it 
may be convenient to step down from that higher value to a 
lower applied potential. 



WO 92/01928 



- 43 - 



PCT/US9J/0507J 



The range of applied potentials at which the 
individual signal measurements may be varied according to 
the needs of the particular application. In the case of a 
glucose sensor, such applied potentials may be chosen to 
lie in the range of about 100 to about 300 mV. Most 
preferably, the one of at least two signal measurements in 
a given time window is carried out at an applied potential 
of about 125 mV and the other measurement at about 225 mV. 
The signal values obtained at each applied potential is 
then plotted on a signal (e.g., current) versus applied 
potential curve and the slope of the line defined by such 
values calculated and compared for each time window. In 
this manner, the slope of the line obtained for a calibrant 
fluid having a known concentration of a preselected species 
may then be compared with that recorded for a line derived 
from a sample fluid. 

As a further illustration of the present 
embodiment, the attention of the reader is directed to 
Figure 4 in which is shown an example of just such a signal 
versus applied potential curve described immediately above. 
On examination of the region of the curve lying between 
about +100 to about +200 mV, one notes that a small net 
positive current can be measured as the electrocatalytic 
oxidation of hydrogen peroxide is not quite counterbalanced 
by the corresponding reduction process. it is clear, 
however, that in this region, or any given region between 
about -100 to about +350 mV, the slope of the curve due to 
plain HEPES buffer (X) is significantly different from that 
of the curve due to a solution of 20 mM glucose in HEPES 
buffer (0). In fact, the slope of each curve is directly 
related to the concentration of glucose ' in each of the 
different fluids. 

Preferably, the applied voltages are centered 
about the region in which the net current is or close to 



WO 92/01928 



- 44 - 



PCT/US91/05071 



zero. In this way, the contribution of interfering redox 
processes to the total current or the effect of any bias 
current associated with measuring electronics is minimized, 
and the best estimate for the maximum slope is obtained. 

5 5.6. SENSOR ACTIVATION 

The mode of operation can be especially complex 
for amperometric sensors which are active devices, unlike 
potentiometric sensors which are electrically passive. An 
amperometric sensor measures the rate of electron transfer 
10 across the electrode-solution interface. Usually the 
electrode surface plays a catalytic role in such electron 
transfer (redox) reactions; therefore, the current is not 
only dependent upon the surface area of the sensor but also 
upon the catalytic activity of the surface. 

15 It is not uncommon for electrode surfaces to be 

contaminated or deactivated. While a catalytic iridium 
surface, acting as the base sensor for a glucose electrode 
( See r for example, prior U.S. Application Serial No. 
07/432,714), is highly active towards hydrogen peroxide 

20 oxidation prior to the deposition of overlaid structures, 
its catalytic activity is much .reduced after such 
processing. The following procedure is designed to recover 
most of the sensitivity of the deactivated metal surface 
without damaging the established overlaid architecture. 

25 5.6.1. METHODS FOR GLUCOSE ACTIVATION 

The present novel operational method is useful 
for rapidly activating the electrode surface of a dry- 
stored amperometric glucose sensor without deleteriously 
affecting the overlaid structures . This activation makes 
30 an important contribution to reducing the overall apparent 
wet -up RC time-constant for the sensors. Unlike the prior 
art pulsing methods, which are always carried out during 
the analytical segment of the measurement (i.e., while the 



WO 92/01928 



- 45 - 



PCT/US91/05071 



devices are fully wet-up and the pulses are related to data 
acquisition), the present activation is carried out in the 
presence of a non-corrosive fluid but prior to data 
acquisition and before the sensor has attained a fully wet- 
5 up state. 

To activate the structure, a set of pulse groups 
is applied to the glucose sensor in the presence of the 
calibrant or sample fluids. This process comprises 
changing (cycling) the applied potential between values of 

10 opposite sign (e.g., +1 V to -1 V). The pulses are applied 
conveniently at full increment steps (i.e., at full ±l v 
steps) . However, it should be apparent to those of 
ordinary skill that such cycling may also be accomplished 
in a variety of other ways including, but not limited to, 

15 (i) pulsing; (ii) intermediate incremental steps to the 
desired positive potential, followed by intermediate 
incremental steps to the desired negative potential; (iii) 
linear potential sweeps to the respective desired 
potentials; and (iv) non-linear potential sweeps or such 

20 sweeps which resemble smooth sinusoidal waves. 

Broadly, the first two pulse groups in the set 
have a magnitude and duration sufficient to activate the 
catalytic surface of the electrode. These pulse groups 
may also promote hydration of the overlaid structures . The 
25 third pulse group in the set serves to reduce the initial 
rate of change in current' upon application of the actual 
operating potential. Referring now to Figure 5a the time 
zero corresponds to the time at which the sensor comes into 
contact with the fluid, calibrant in the preferred 
30 instance), one observes that during the application of the 
first two pulse groups, the peak currents for hydrogen 
peroxide oxidation and reduction increase dramatically. 
Figure 5b shows in more detail the response of the 
activated sensor, at the operating potential, to a 5 mM 
35 glucose calibrant solution (26-53 sec) followed by a 10 mM 



WO 92/01928 



- 46 - 



PCT/US91/05071 



human serum (from about 54 sec) . The analytical value of 
the glucose sensor is clearly apparent, with a significant 
current being observed for both fluids and in which the 
current shows an increase after the sensor is exposed to a 
5 biological fluid having a glucose concentration which is 
higher than the calibrant solution. At an operating 
potential of +350 mV, the sensor gives a linear current 
response at a range of 0,1-30 mM glucose. Figure 5c, 
suggests that the sensor is almost fully activated because 
10 there is only a marginal increase in the maximum hydrogen 
peroxide oxidation and reduction currents when the entire 
pulsing sequence is repeated on the same sensor. 

A viable alternative to cycling the applied 
potential, involves galvanostatic control over the sensor 
15 activation with a constant current source being applied to 
the sensor until the required potential attains some 
predetermined value or rate of change. 

Another alternative for measuring the current 
output of the sensors, to determine its level of 
20 activation, involves making a conductivity measurement 
between two electrodes, one which is the sensor. 

In one embodiment of the present method the 
pulsing sequence for hydrogen peroxide measuring sensors 
are as follows : pulse groups 1 and 2 should be of a 

25 duration of about 5 seconds with limits Of +0.7 to +1.2 V 
at the oxidizing end and -0.7 to -1.2 V at the reducing 
end, which pulses may or may not be applied symmetrically. 
A single .extended negative pulse followed by the standard 
sequence may also be applied advantageously. The inventors 

30 have observed that negative pulses are an important aspect 
of the activation process, and they speculate that the 
activation process is likely to be associated with the 
reduction of one or more types of functional groups of the 
iridium electrode. However, the inventors do not wish to 



WO 92/01928 



- 47 - 



PCT/US91/05071 



be bound by the present speculation nor do they wish to 
limit the scope of the present invention by making specific 
interpretations of the surface activation process on a 
molecular level. 

5 The third P ulse group should be selected from 

groups in which the final potential step is in the range 
+400 mv to +800 mV. As can be seen from Figure 5b the 
current recorded for the 5 mM glucose calibrant fluid shows 
that the sensor has attained a sufficient degree of wet-up 
10 and activation. During the period from 46 to 52 seconds, 
data is recorded in the first time window for later 
extrapolation. After 53 seconds the calibrant fluid is 
forced to waste and the biological fluid (sample) placed 
over the sensor. An advantage of thin-film microfabricated 
biolayers is that the response-time (i.e., the time it 
takes for the current to change to a value which is 
proportional to the. bulk concentration of the analyte in 
the second fluid) is fast, usually less than 5 seconds. It 
is this additional property of the present microfabricated 
devices that contributes to the success of the present 
analytical methods. During the period from 61 to 66 
seconds, the data is recorded in the second time window. 

As with the potentiometric sensors, the 
selection of an appropriate location for the data 
25 acquisition window and the choice of the data fitting 
computational method are based on empirical calculations of 
the accuracy and precision attainable with each of the 
fits. As with the potentiometric urea sensor a Linear/Mean 
fit is preferred for the glucose sensor. 



15 



20 



30 



5.6.2. OPERATIONAL METHODS GENERALLY 

The discussion above relates to a glucose sensor 
comprising an iridium electrocatalyst at which hydrogen 
peroxide produced by an enzymatic reaction is measured by 



WO 92/01928 



- 48 - 



PCT/US91/05071 



means of electrochemical oxidation at an operating 
potential of +350 mV versus an on-board silver-silver 
chloride reference electrode. An alternative method for 
operating this sensor is based upon electrochemical 
5 reduction of hydrogen peroxide. This operation is achieved 
by applying an operating potential in the range of zero mV 
to -250 mV, preceded by pulse groups similar to 1 and 2 
above, but with pulse group 3 having a final potential set. 
in the range of -200 mV to -500 mV. 

10 The operational methods described above can also 

be applied for activating sensors where platinum or another 
noble metal is used as the electrocatalyst in place of 
iridium; these surfaces also become deactivated during 
post-processing steps* 

15 An oxygen sensor of the type disclosed in the U. 

S. Application Serial No. 07/432,714, with a gold 
electrocatalyst suitable for oxygen reduction, may also be 
operated with this type of activation method. However, it 
is sometimes desirable to modify the duration and magnitude 

20 of pulse groups 1 and 2 because different metal surfaces 
are deactivated to different extents during the deposition 
of overlaid structures and related processing steps. In 
addition, pulse group 3 is preferably chosen to fall in the 
range of -600 mV to -800 mV, where an operating potential 

25 in the range of -400 mV to -550 mV is contemplated for 
oxygen reduction at a gold electrode. 

5.6.3. COMPUTATIONAL METHODS FOR 

ENZYME ASSAYS AND IMMUNOASSAYS 

In a metabolite assay the object of the 
30 computational method is the determination of the bulk 
concentration of the metabolite. Clearly, the sensor 
should not perturb the bulk concentration of the metabolite 
if the measurement is to be of analytical value. 



WO 92/01928 



- 49 



PCT/US91/05071 



10 



For an enzyme sensor or an enzyme-linked 
immunosensor-based assay, by contrast, the object of the 
sensor computational method is the determination of the 
rate of change in concentration of an electroactive species 
which is consumed or produced while the assay is in 
progress, which rate reflects the enzyme activity present 
in the system. Sensors appropriate for enzyme and enzyme- 
linked immunoassays are disclosed once again in U.S. 
Application Serial No. 07/432,714. 



If the product of the assay is hydrogen peroxide 
(i.e., it is the electroactive species which is produced in 
the course of the assay), the computational method can be 
modified as follows: after a set of activation pulse groups 
is applied and measurement of a calibrant current is 
15 performed, as described above, substrates or reagents for 
the enzyme or enzyme-linked immunoassay are introduced to 
the sensor. After a brief induction or mixing period the 
current changes steadily, and the initial rate of change, 
di/dt, is then computed (Sea, Figure 6) . From the rate of 
20 change in the current and the known current electroactive 
species corresponding to a certain concentration in the 
calibrant, the rate of increase in it's concentration can be 
estimated. This rate can be expressed as the enzyme 
activity, that is moles of substrate consumed per unit time 
25 at a certain temperature, pH etc. It is important to note 
that if the data collection period extends over several 
minutes, the above-mentioned pulse groups wet-up the sensor 
structure to a sufficient extent such that the calibrant 
signal can be readily extrapolated forward in the time 
30 domain. Otherwise the rate measurement, which is based on 
the initial sensor response in the presence of calibrant 
fluid (if calibrant is introduced prior to introducing the 
sample), may be severely compromised (Refer to earlier 
discussions related to the importance of making the 
calibrant and sample measurements close in time) . The use 



35 



WO 92/01928 



- 50 - 



PCT/US91/05071 



of microfabricated immunosensors in this manner makes it 
possible to obtain rate measurements prior to the sensor 
attaining a fully wet-up state. 

Another problem that is encountered when a 
5 sensor is operated over an extended period (2-3 minutes) is 
that product build-up or reaction depletion zones occur in 
the overlaid structures of the electrode with concomitant 
diffusion gradients extending out into the bulk solution. 
Such inhomogeneity close to the sensor can adversely affect 

10 its performance. For example, an excessive differential 
between the hydrogen ion concentration close to the 
electrode and hydrogen ion concentration in the bulk 
solution causes a shift in the activation energy for 
hydrogen peroxide oxidation. This shift may result in a 

15 non-linear current response. This problem may be 
circumvented by modifying the computational method to 
include additional sets of pulse groups at preselected time 
intervals. After applying the initial set of three pulse 
groups and the calibration has been performed, as described 

20 above, substrates or reagents for the assay are allowed to 
pass over the sensor* A measurement is then made in much 
the same vay, as described above for glucose. However, 
after this measurement the pulsing sequence, or some 
modified portion thereof, is repeated and a second 

25 measurement is made. This sequence is repeated several 
times (usually five to ten) to yield a .set of current 
measurements made at exact time intervals; The initial 
rate of change in the current may then be more accurately 
estimated from this set of current measurements because 

30 repeated pulsing has the effect of washing-out the 
electrode surface and overlaid structures and disordering 
the fluid layer and concentration gradients which lie close 
to the sensor. Again all of these processes are monitored 
on a sensor that is undergoing wet -up, i.e., t ^ 3T. 



WO 92/01928 



- 51 - 



PCT/US91/05071 



5.6.4. ADVANCED OPERATIONAL METHODS 

Advanced operational methods can also be used to 
make the system "smart." Such methods can be applied 
during the real time operation of the sensors or during 
5 post data collection treatments. In particular, the 
flexible methods for governing the activation of 
amperometric sensors, assessing the propriety of a given 
set of data points or determining the best position for the 
data window may be incorporated, to name a few. 

10 Thus, these methods may include alternatives to 

applying the identical pulse sequence to all amperometric 
sensors which may be present in the array. For example, 
the operational method can be modified such that pulse 
groups 1 and 2 are applied until the observed peak current 

15 associated with each potential pulse has reached a finite 
value, until the charge passed in each pulse has reached a 
finite value, until the charge passed for the entire pulse 
sequence has reached the finite value or until the RC time 
constant for the sensors after a given pulse is within some 

20 predetermined range of values. Moreover, pulse groups 1 
and 2 may be applied until the rate of change between 
successive pulses, in terms of peak current, charge passed 
or RC time constant, is within some preselected range. 

Advanced methods for error recovery can also be 
25 implemented. In particular, computational methods can be 
put in place for detecting the presence of glitches in the 
data set, whether such glitches are isolated or numerous, 
or whether they may be associated with common background 
noise. An advanced method for sensing integrity failures 
30 may suggest appropriate solutions, such as the application 
of median filters. Alternatively, another data set may be 
chosen which is derived from a separate data window stored 
in the system. For example, if the sample causes an 



WO 92/01928 



- 52 - 



PCT/US91/05071 



abnormally slow response from the sensor, gives rise to 
non-ideal waveforms upon contact with the sensors or, 
generally, does not provide well-behaved waveforms, the 
first and second time windows may not be positioned in the 
5 best place initially. The presence of slight defects in 
the the sensor may also, give rise to less than desirable 
responses. In the case of a slow response, the maximum 
value for the sample or second fluid may not have yet been 
reached before a second signal measurement is performed 

10 (i.e., a second time window is selected for comparison with 
the first time window associated with the first or 
calibrant fluid). The signal associated with the sample 
may degrade rapidly soon after reaching the maximum value, 
thus providing a lower value than what should have been 

15 observed if the signal had not degraded. 

In both cases discussed in the preceding 
paragraph, the computational method may search for the 
proper data collection window and locate the maximum value 
whiclx provides a better estimate of the concentration of 

20 the preselected analyte species. Also, even if certain 
error limits are reached or exceeded, a reinspection of the 
value of the slope and the quality of the overall drift may 
prompt the instrument to accept the values obtained. In 
certain cases, the value of the delta drift can override a 

25 data window with an unusual number of glitches or a high 
level of noise. Of course, different methods for 
extrapolating the slopes and calculating the difference 
between the signal measurements may be selected. These 
non-ideal responses may be encountered more frequently with 

30 sensors of increased functional complexity. 

As mentioned earlier, one may also conceive of 
flexible computational methods adapted to determining when 
and whether a sufficient number of data points has been 
collected. Such advanced methods may decrease the time 



WO 92/01928 



- 53 - 



PCT/US91/05071 



10 



20 



25 



30 



necessary for carrying out the data collection and 
analysis. 

5.7. CONDUCTIVITY MEASUREMENTS 

For a colloidal suspension such as blood, the 
electrical conductance is a function of the nature of the 
particles, the medium and the fraction of the total volume 
occupied by the particles. At high particle 

concentrations, as in blood, it is also necessary to take 
into account the geometry and orientation of the particle. 



The entire period of data acquisition in the 
calibrant and in the blood takes only about one second. 
During this period the a.c. conductivity is measured at a 
frequency which is selected to be sufficiently high to 
minimize the impedance at the sensor-fluid interface, and 
15 sufficiently low to minimize capacitive coupling across the 
erythrocyte cell membrane. A preferred frequency is about 
50 KHz . 



Once the calibration has been made, the 
calibrant fluid is then removed and blood introduced. The 
measurement cycle is then repeated with the actual 
measurement time being selected so that the blood is in a 
quiescent state, but before a significant degree of 
erythrocytes have settled. 

The conductivity sensor comprises two noble 
metal electrodes microfabricated on a planar surface and 
designed with the appropriate geometry. 

For conductivity measurements there is no need 
for a sophisticated computational method for performing a 
signature analysis. Calculation of the percent hematocrit 
in the blood is made either from an empirically determined 
calibration curve stored in the electronics or, more 
preferably by means of an equation developed by Velick and 



WO 92/01928 



- 54 - 



PCI7US91/05071 



Gorin as described in Journal Ql General Physiology 1940, 
21, 752-771. 

Such conductivity measurement may also be 
utilized to determine the success or failure of the fluid 
5 change operation. That is , stored values of predicted 
fluid conductivities can be compared to the observed 
measurements to provide a quality assurance method for 
detecting the presence of an adequate intervening air 
segment between fluids, the failure of the calibrant fluid 
10 to move to waste, or other such system failures associated 
with the fluidics movements. 

5.8. SYSTEM INTEGRATION 

As alluded to elsewhere in this disclosure, a 
system most attractive in the clinical setting is not 

15 limited to discrete measurements of single analytes. 
Instead, an array of sensors, designed to make a 
multiplicity of discrete measurements of a range of 
different analytes in biological fluids is preferred. This 
array of sensors is preferably exposed to a single, common 

20 calibrant. fluid which is removed after all of the sensors 
have been calibrated. Only then can the second (sample) 
fluid be introduced to the sensors. Such an integrated 
setting means that all of the sensors must be chemically 
and electrically compatible: that is, they must wet-up and 

25 respond at approximately the same rate without interfering 
with one another. Thus, sensor compatibility and system 
integration is enhanced if the pulsing sequence, used in 
amperometric measurements, is completed prior to 
performing potentiometric measurements because the high 

30 current flow in solution associated with pulsing can 
undermine the integrity of the potentiometric signal. 

Also, when the conductivity sensor is activated, 
a current of ca. 10~ 3 A, flows in the solution between the 



WO 92/01928 



- 55 - 



PCT/US91/05071 



15 



pair of metal electrodes. While this current may seem 
small, it is about six orders of magnitude greater then 
that measured in the amperometric based analyses. Again, 
during the period when the amperometric sensors are being 
5 pulsed, a maximum current of, ca. 10-6 A, also flows. To 
minimize interference during calibration between the 
conductimetric and amperometric sensors, it is preferable 
that the conductivity sensor be operated after activation 
of the amperometric sensors but before amperometric data 
10 are acquired. After the fluid change, the amperometric 
data acquisition is performed before the second 
conductivity measurement is made. A series resistor, ca. 
105 ohm, may also be used to protect amperometric sensors 
during the a.c. conductivity measurements. 

As mentioned previously, the conductivity sensor 
may also be used to distinguish the general composition of 
a fluid, i.e., whether the sensor is in contact with 
calibrant fluid, serum or whole-blood, or for that matter 
no sample at all, i.e., air. Because many of the sensors 
display matrix effects, as discussed previously above, this 
measurement may be used to make the appropriate correction 
to the calculation of the concentration of the preselected 
analytes. 

In terms of the integration of conductivity 
25 measurements with potent iometric measurements, there is no 
discernable interference. However, it is preferred that 
the conductivity measurement be avoided during data 
acquisition at the potentiometric sensors. 

In addition, it is preferred that there exists a 
fluid grounding electrode in close proximity to the 
potentiometric sensors to absorb excess charge generated in 
the fluid when the conductivity sensor is activated. This 
excess charge may have the undesirable effect of polarizing 
the potentiometric membranes. Moreover, the membranes must 



20 



30 



WO 92/01928 



- 56 - 



PCT/US91/05071 



be depolarized rapidly if they are to be of analytical 
value. 

Finally, it may be desirable to place the 
conductivity measuring sensor on a separate chip from that 
5 of the potent iometric sensors. Commonly, these sensor 
structures are actually fabricated on a silicon wafer with 
an insulating layer, ca. 0.5 1CT 6 m in thickness, of 
silicon dioxide . The underlying silicon is a semiconductor 
which means that capacitive coupling between sensors 
10 presents a possible signal interference. Separating the 
sensors on different chips obviates this potential problem. 

6 . EXAMPLES 

As a further illustration of the present method, 
the following example is described. Referring now to Chart 

15 II, there is shown a more detailed flow chart of the Data 
Collection method.; As mentioned previously, data 
collection may be accomplished in several different ways, 
including simply sequentially measuring and storing data. 
The method of the preferred embodiment permits verifying 

20 integrity of the measurements in real time, while 
collecting input data. Although a thorough data integrity 
analysis is provided at the Data Analysis stage, this on- 
line verification method permits early detection and, 
possibly, elimination of spikes, glitches, and other noise 

25 in the data. Alternatively, time windows can be determined 
intelligently at flexible intervals where the data points 
are not significantly affected by noise. Such smooth time 
windows can thus be located readily. 



- 57 - 



PCI7US9I/05071 



CHART II 

21 

" \ COLLECT DATA POINT | — 

22 [ 

INCREMENT DATA COUNTER 



DATA COUNTER 
£ SLIDING POINTS 



26 



COMPUTE RATIO 07 
SLIDING WINDOW SLOPE 
TO BASIS WINDOW SLOPE 



27 



SET 


ERROR BIT IF 


(RATIO 


> UPPER BOUND) 




AND 


(RATIO 


< LOWER BOUND) 




COMPUTE 


BASIS 


WINDOW 


SLOPE 





WO 92/01928 



- 58 - 



PCT/US91/05071 



In the present example, the Data Collection 
process begins at block 21 , where each data point is 
measured. Data acquisition involves taking an analog 
signal and recording it in digital format in the 
5 appropriate memory location. To ensure that the system is 
not susceptible to 60 cycle noise, a set of 80 data points 
every one-sixtieth of a second are preferably recorded. 
(Clearly in countries wh£ re 50 cycle noise is the norm the 
sampling rate is set at 80 data points every one-fiftieth 
10 of a second) . The average value of this set is then 
calculated, and this value is taken as a "single" data 
point. These "single" data points are then used in further 
calculations. 

As described above, the electronics are designed 
15 to be appropriate for high impedance potentiometric sensors 
with sufficient resolution over the expected range of 
measurements. The' electronics for amperometric sensors 
includes current to voltage convertors designed to have 
sufficient resolution over the expected range of current 
20 measurements. 

Next, the method flow passes to block 22, where 
a counter (DATA_COUNTER) , which keeps track of the 
aggregate number of data points collected within a time 
window, is incremented by one, (The DATA_COUNTER is 

25 initialized to zero prior to acquisition of the first data 
point of a given time window.) The method flow then passes 
to test 23, where the total number of collected data points 
(DATA COUNTER) is compared to the number of data points of 
a selected sliding window set (SLIDING_POINTS) . The size 

30 of a sliding window is selected on the basis of a trade-off 
between the computational time and precision of error 
detection. In this instance, a seven point sliding vindow 
is selected from a twenty-five point time window. 
According to test 23, if the total number of acquired 



WO 92/01928 



- 59 - 



PCT/US91/05071 



points, is less than the size of one sliding window, the 
control returns to block 21, where another data point is 
acquired. Otherwise, control is transferred to test 24. 

Test 24 is provided to locate the basis window. 
5 In this instance, the basis window is defined as the first 
sliding window, and remains the same throughout the data 
acquisition process for a given time window. it should be 
apparent to one of ordinary skill in the art that there are 
other methods of selecting a basis window. For example, a 
10 basis window can be defined as a sliding window which is 
"behind" the current sliding window by one or more points. 

According to test 24, if an exact number of 
SLIDING_POINTS are collected, method flow is transferred to 
block 29; otherwise, the total number of collected points 

15 is greater than the size of one sliding window, and method 
flow passes to block 25. At block 29, the slope of the 
basis window is computed using a linear regression (i.e., 
by fitting points into a line defined by y = ax + b, 
wherein the slope is "a"). At block 25, a new sliding 

20 window set is formed and a recursive form of the linear 
regression method is used to compute the slope of a sliding 
window. m this instance, a new sliding window set is 
created whenever flow enters block 25. Therefore, a first 
sliding window consists of the following data points: {2, 3 
25 ... (SLIDING_POINTS + 1)), and each subsequent sliding 
window is formed by including a newly acquired data point 
and eliminating the first data point of the previous 
sliding window set. 

Next, the method flow passes to block 26, where 
30 the slope of the basis window is compared to the slope of 
the current sliding window, and the ratio of the slopes is 
computed. Then at block 27, the integrity of a current 
sliding window data set is verified. If the ratio of the 
slopes is not within acceptable bounds, the output of such 



WO 92/01928 



- 60 - 



PCT/US91/05071 



sensor is deemed to contain an error. One can set, for 
example, an upper bound of 2.5 for the ratio. If, however, 
the basis window has a value of zero, a difference instead 
of a ratio is used. 

5 Hence, if the ratio or difference is out of 

bounds, a flag (ERROR) is set. In this instance, the ERROR 
flag is simply a bit. Alternatively, the ERROR flag value 
might contain a pointer to a specific data point that 
causes the ratio of the slopes to be out of bounds. Next, 

10 the method flow enters test 28, which checks whether 
additional data points should be measured for a given time 
window. If all the time window points <WJPOINTS) have not 
been collected, the flow returns to block 41; otherwise, 
the control is transferred to a new process step which 

15 causes the displacement of the calibrant fluid by the 
sample fluid, with an air segment present between the 
fluids. When the last time window has been collected, the 
following data analysis stage is carried out. It should be 
apparent that W_POINTS are collected for each active sensor 

20 in the array. Thus, each process block of Chart II is 
performed essentially in parallel for each active sensor. 

" Chart III shows schematically the steps of the 
Data Analysis stage of the method in which analyte 
concentrations are determined from each sensor measurement. 
25 At this stage, data points for the first (e.g., calibrant) 
and second (e.g., sample) time windows have been collected 
and stored in appropriate memory locations. Test blocks 
41-47 occur for each active sensor in the array, while test 
blocks 48-51 occur once for the entire array. 

30 The process begins at test 41 which checks 

whether an error has been detected at the Data Acquisition 
stage by testing the status of the ERROR bit. If the ERROR 
bit is set, method flow passes directly to block 42, where 



WO 92/01928 



- 61 - 



PCT/US91/05071 



error recovery is performed. Otherwise (i.e., no errors at 
Data Acquisition stage), flow passes to block 42. 



- 62 - 



PCT/US91/05071 




CHART III 



DETERMINE SLOPES IN 
THE TIME WINDOWS 



NO ERRORS 



47 




EXTRACT DELTA 
RESPONSE 



45 



MARK SENSOR 
FAILED 



ERROR RECOVERY 



48 



ADJUSTMENTS DUE TO 
REF ELECTRODE RESPONSE 



49 



CALCULATE 
CONCENTRATION 



50 



STANDARDIZE 



51 



DISPLAY CONCENTRATIONS 
OR ERROR MESSAGES 



WO 92/01928 



- 63 - 



PCT/US91/0507! 



At block 42, a linear regression is applied to 
the data points of each window to approximate the drift 
rates (slopes) of the data window. A linear least squares 
fit is a well known method of interpolation, in which data 
5 points are connected to approximate a line, w y = ax + b, " 
where constants n a n and "b" are chosen such that the sum of 
squares of the deviation from the actual data points is 
minimized. Alternatively, higher order regressions, i.e., 
those that approximate 2nd, 3rd, nth order functions, 

10 can be implemented easily. Generally, the order of the 
applied regression should be determined according to the 
nature of the sensors' degree of wet-up and waveform in the 
selected time window intervals. 

Next, the flow passes to test 43, where the data 
15 points, of both windows are checked for unexpected values. 
The following data integrity verification is then 
performed: 

MIN_SLOPE < first window slope < MAX_SLOPE • 
MIN_SL0PE < second window slope < MAX_SLOPE 
20 (first window slope - second window slope) < MAX_DIFF 

MIN_MEAN < first window mean -value < MAX_MEAN 
MIN_MEAN < second window mean value < MAX_MEAN_1 
first window error of approximation < MAX_ERR0R 
second window error of approximation < MAX_ERROR 

25 The values, MIN_SL0PE, MAX_SLOPE, MAX_DIFF, 

MIN__MEAN, MAX_MEAN_1 and MAX_ERROR are expected ranges of 
the above parameters. The specific values are determined 
experimentally, based on the predictable characteristics of 
the sensors' response curve. Table I lists some suggested 

30 values for the various data window parameters. 



WO 92/01928 



- 64 - 



PCT/US91/05071 



TABLE I 

SUGGESTED LIMITS FOR VARIOUS DATA WINDOW PARAMETERS 
. FIRST WINDOW 





SLOPE 


(mV/s) 


ME AM 


(mV) 


ERROR (mV) 




min 


max 


min 


max 


max 


CI 


-0.65 


0.65 


-200 


+100 


0.2 


K 


-0.25 


1.0 


-250 


+100 


0.2 




0.0 


1.0 


-250 


+100 


0.2 


BUN a 


-0.65 


0.65 


-200 


+100 


0.2 


Glue 15 


-0.02 c 


0.02 c 


0.2 d 


3d 


0.015 d 








SECOND 


.WINDOW 






SLOPE 


(mWs) 


mean, 


, (mV). 


ERROR (mY> 


SENSOR 


min 


max 


min 


max 


max. 


CI 


-.0.65 


0.65 


-200 


+100 


0.2 


K 


-0.25 


1.0 


-250 


+100 


0.2 


Ma 


0.0 


1.0 


-250 


+100 


0.2 


BUN a 


-1.0 


1.0 


-200 


+100 


0.4 


Gluc*> 


-0.1 c 


0.1 c 


0.05 d 


32 d 


0.015 d 



20 a CI, K, Na and BUN stands for chloride, potassium, sodium 
and Blood Urea Nitrogen sensors, respectively. b Gluc 
stands for Glucose sensor. c This value is in units of 
nA/s. d This value is in nA. 



WO 92/01928 



- 65 - 



PCT/US91/0507I 



In the present instance, the limit on the 
difference (in mV/s) between the value of the slope in each 
time window is set at a maximum of 0.50, 0.30, 0.30 and 
1.00 for the chloride, potassium, sodium and BUN sensors, 
5 respectively. The observed waveform for the second fluid, 
such as glucose, in contact with the microf abricated 
amperometric sensors can exhibit a maximum or minimum' 
value; consequently a linear/mean fit is preferred. 

If one of the above tests indicates that the 
10 measurement is out of range, all the measurements generated 
by a particular sensor are discarded as unreliable. 
However, the present method provides for error recovery of 
corrupt data. Accordingly, if an error is detected at test 
43, the method flow passes to test 44. If no errors are 
15 detected, control is transferred to block 47, where the 
sensors 1 responses are determined. 

Test 44 determines whether a previous attempt to 
correct errors has already been made. At this step, a 
variable "FIRSTJTIME" is incremented by one. (FIRST_TIME 

20 is initialized to zero at the beginning of the Data 
Analysis for each time window.) A value of FIRSTJTIME that 
is greater than 1 indicates that a previous attempt to 
correct errors has been made. Because the flow returned 
subsequently to test 44, error recovery failed once and 

25 should not be repeated. In this case, method flow passes 
to block 46, where the "FAILED" bit is set to indicate that 
this particular measurement has failed, and that the only 
further processing remaining for such sensor involves 
displaying an error message at block 51. Otherwise, the 

30 value of FIRST_TIME is one and the flow passes to block 46, 
where error recovery is attempted. 

Error recovery can be implemented in a variety 
of ways. For example, noise and spikes in the data can be 



WO 92/01928 



- 66 - 



PCT/US91/05071 



eliminated by known digital techniques, such as median 
filters and the like described previously, i.e., the re- 
evaluation of observed values e.g., p. 29.) Another 
method may involve the interpolation of a curve on the 
5 basis of the first several points of a given time window 
and the rejection of those points that deviate 
significantly from .the interpolation. In most cases, it 
may be desirable to collect more data points than that 
required for a given time window, so that error recovery 
10 may be accomplished by selecting an alternative time window 
set. From error recovery at block 4 6, method flow turns 
back to block 42 where the points of the corrected data set 
are interpolated and new slopes and errors of approximation 
are computed. 

15 if collected data passes all the integrity 

tests, the control is transferred to block 47 where the 
sensors 1 responses are determined by relating measurements 
in the first (calibrant) time window to those in the second 
(sample) time window. For potent iometric sensors, the 

20 analytical value of interest is the delta response, which 
is the difference between the amplitudes measured at a 
selected point of an analyte and calibrant response curves . 
For amperometric sensors, the delta response corresponds to 
the ratio of the respective amplitudes • As mentioned 

25 above, the data acquisition is performed while the sensors 
exhibit a monotonic wet-up response. In this instance, the 
responses are measured in time windows which are selected 
such that linear interpolations are sufficiently accurate 
to describe the response drifts. Consequently, linear 

30 approximations of block 42 can be reliably projected 
forward and backward in time to compare each sensor's 
response to the fluid change, AS mentioned previously, nth 
order approximations can also be used for this purpose. 

The particular sensor's delta response is then 
35 calculated in one of the many different ways including, but 



WO 92/01928 



67 - 



PCI7US91/05071 



not limited to, a Linear/Linear delta, Linear/Mean delta, 
Mean/Mean delta, Mean/Linear delta, Maximum/Linear delta, 
Linear/Maximum delta, Linear /Minimum delta and 
Minimum/Linear delta approaches. In the Linear/Linear 
5 case, the fit to the first window is extrapolated forward 
to an estimated fluid transition point, i.e., the midpoint 
between the first window and the second window. The fit to 
the second window (sample) is extrapolated backward to the 
same estimated transition point. The difference between 
10 these extrapolations is the particular sensor's delta 
response. In the Linear/Mean approach, the linear fit to 
the calibrant is extrapolated forward to the midpoint value 
of the sample window and compared to the midpoint value of 
the sample response curve. In the Mean/Linear approach the 
15 previous sequence is reversed. In the Mean/Mean approach, 
the delta response is the difference between the midpoint 
of the sample window and the midpoint value of the 
calibrant window. ; The Linear /Maximum, Maximum/Linear, 
Linear/Minimum and Minimum/Linear methods are analogous to 
20 Linear /Mean and Mean/Linear methods, except that the mean 
value is replaced by the maximum or minimum value of the 
corresponding time window. On the basis of routine 
experimentation, one of the above methods can be found more 
desirable than the others for obtaining a more accurate 
25 measure of the concentration of a particular analyte when 
compared to a reference method of analysis. Presently, the 
Linear/Linear method is found to be superior for the 
potassium, sodium and chloride sensors. A Linear /Mean 
method is best for the present embodiment of the glucose 
30 and urea sensors. 

^ After the delta responses are computed for each 
sensor, the flow passes to block 48. At block 48, the 
response of the reference electrode is subtracted from the 
delta response determined in block 47. At this stage, 
35 certain corrections are made for a slight response of the 



WO 92/01928 



- 68 - 



PCT/US91/05071 



reference electrode due to matrix effects and differences 
in ionic strength of the fluids. Next, the flow passes to 
block 49, where, for potentiometric sensors, Equation 1 and 
coefficients derived from Equation 2, as described above, 
5 are applied to determine analyte concentrations. For 
amperometric sensors, Equation 3 is applied. Then, at 
block 50, the results are appropriately scaled in order to 
derive standardized values. Finally, block 51 provides a 
display of the calculated concentrations or a display of an 
10 appropriate message if aberrant values are found. 

The preceding example is presented solely to 
illustrate a method for practicing the invention and should 
not be construed as limiting the invention in any way. 
Doubtless, other embodiments may be conceived which would 
15 not depart significantly from the spirit and of the present 
invention, which scope is defined by the following claims. 



WO 92/01928 



69 - 



PCT/US9I/05071 



10 



WHAT IS CLAIMED IS: 

1. A method of determining the concentration 
of a preselected analyte species present in a sample fluid 
comprising: 

(a) providing an external computational means, 
a reference electrode and at least one substantially dry- 
Stored sensor capable of exhibiting a response to changes, 
in the concentration of a preselected analyte species 
before said sensor attains full equilibrated wet-up; 

(b) establishing electrical contact between 
said sensor, reference electrode and external computational 
means; . 

(c) contacting said sensor and reference 
electrode with a calibrant fluid; 

15 < d > performing a first signal measurement in a 

first time window in the presence of said calibrant fluid; 

(e) displacing said calibrant fluid after 
performing said first signal measurement; 

(f) contacting said sensor and reference 
20 electrode with a sample fluid; 

(g) performing a second signal measurement in a 
second time window in the presence of said sample fluid; 
and 

(h) relating said first and second signal 
25 measurements to determine the concentration of said 

preselected analyte species in said sample fluid before 
said sensor attains full equilibrated wet -up. 



WO 92/01928 



- 70 - 



PCT/US91/05071 



2. A method of determining the concentration 
of a preselected analyte species present in a sample fluid 
comprising: 

(a) providing at least one microf abricated 
5 potentiometric sensor sensitive to changes in the 
concentration of a preselected analyte species , a reference 
electrode capable of sustaining a well-behaved reference 
potential for a sufficient period of time and external 
computational means, 

!0 which sensor and reference electrode have been 

stored substantially dry and which sensor exhibits a rapid 
response to said changes in the concentration of said 
preselected analyte species before said sensor attains full 
equilibrated wet-up; 

!5 (b) establishing electrical contact between 

said sensor, reference electrode and external computational 
means; 

(c) contacting said sensor and reference 
electrode with a calibrant fluid; 

20 (d) performing a first signal measurement in a 

first time window in the presence of said calibrant fluid; 

(e) displacing said calibrant fluid after 
performing said first signal measurement; 

(f) contacting said sensor and said reference 
25 electrode with a sample fluid; 

(g) performing a second signal measurement in a 
second time window in the presence of said sample fluid; 
and 



WO 92/01928 



- 71 - 



PCT/US91/05071 



.(h) relating said first and second signal 
measurements to determine the concentration of said 
preselected analyte species in said sample fluid before 
said sensor attains full equilibrated wet-up. 

5 3. A method of determining the concentration 

of a plurality of different preselected analyte species 
present in a sample fluid comprising: 

(a) providing an array of raicrof abricated 
potent iometric and amperometric sensors, each sensitive to 

10 changes in the concentration of different preselected 
analyte species, and external computational means, 

which array comprises individual sensors and at 
least two reference electrodes capable of sustaining a 
well-behaved reference potential for a sufficient period of 

15 time, one reference electrode being connected to said 
potentiometric sensors when the other reference electrode 
is connected to said amperometric sensors, and each of 
which sensors has been stored substantially dry and 
exhibits a response to said changes in the concentration of 

20 the different preselected analyte species before each 
attains full equilibrated wet -up; 

(b) establishing electrical contact between 
said array of sensors and external computational means; 

(c) contacting said array of sensors with a 
25 calibrant fluid; 

(d) performing a first set of signal 
measurements in a first time window in the presence of said 
calibrant fluid; 

(e) displacing said calibrant fluid after 
30 performing said first set of signal measurements; 



WO 92/01928 



- 72 - 



PCIYUS91/05071 



(f) contacting said array of sensors with a 
sample fluid; 

(g) performing a second set of signal 
measurements in a second time window in the presence of 

5 said sample fluid; and 

(h) relating said first and second sets of 
signal measurements to determine the concentration of a 
plurality of the different preselected analyte species in 
said sample fluid before each sensor attains full 

10 equilibrated wet-up. 

4 . A method of determining the ratio of the 
concentrations of a preselected analyte species present in 
two fluids comprising: 

(a) providing at least one microf abricated 
15 amperometric sensor sensitive to changes in the 
concentration of a preselected analyte species, a reference 
electrode capable of sustaining a well-behaved reference 
potential for a sufficient period of time and external 
computational means, 

20 which sensor and reference electrode have been 

stored substantially dry and which sensor exhibits a rapid 
response to said changes in the concentration of said 
preselected analyte species before said sensor attains full 
equilibrated wet-up; 

25 (b) establishing electrical contact between 

said sensor, reference electrode and external computational 
means; 

(c) contacting said sensor and reference 
electrode with a first fluid; 



WO 92/01928 



- 73 - 



PCT/US91/05071 



5 



(d) performing a first signal measurement in a 
first time window in the presence of said first fluid; 

(e) displacing said first fluid after 
performing said first signal measurement; 

(f) contacting said sensor and said reference 
electrode with a second fluid; 

(g) performing a second signal measurement in a 
second time window in the presence of said second fluid; 
and 

10 (h) relating said first and second signal 

measurements to determine the ratio of the concentrations 
of said preselected analyte species in said first and 
second fluids before said sensor attains full equilibrated 
wet -up. 

15 5 • A method of determining the ratio of the 

concentrations of a preselected analyte species present in 
two fluids comprising: 

(a) providing at least one dry-stored 
microfabricated sensor capable of exhibiting a response to 
changes in the concentration of a preselected analyte 
species, which response is fast relative to the monotonia 
wet-up behavior of said sensor when contacted with a fluid; 

(b) providing a dry-stored microfabricated 
reference electrode capable of sustaining a well behaved 

25 reference potential for a period of time sufficient to 
permit the completion of at least two signal measurements 
and exhibiting a monotonic wet-up behavior which is similar 
to that exhibited by said sensor when contacted with a 
fluid; 



20 



WO 92/01928 



- 74 - 



PCT/US91/05071 



(c) establishing electrical contact between 
said sensor, reference electrode and external computational 
means ; 

(d) contacting said sensor and reference 
5 electrode with a first fluid; 

(e) performing a first signal measurement in a 
preselected first time window in the presence of said first 
fluid before said sensor and reference electrode attain 
full equilibrated wet-up; 

10 (f) displacing said first fluid after 

performing said first signal measurement; 

(g) contacting said sensor and reference 
electrode with a second fluid; 

(h) performing a second signal measurement in a 
15 preselected second time window in the presence of said 

second fluid before said sensor and reference electrode 
attain full equilibrated wet-up; 

" (i) relating the first and second signal 
measurements to determine the ratio of the concentrations 
20 of said analyte species in said first and second fluids by 
a computational method which distinguishes the relatively 
fast response of the sensor to changes in the concentration 
of said preselected analyte from the slower monotonic wet- 
-up behavior of said sensor and reference electrode. 

6. The method of claim 1, 2 or 3 in which the 
order by which said sensor, reference electrode or array 
thereof is contacted with said calibrant and sample fluids 
is reversed. 

7. The method of claim 6 in which said first 
30 and second signal measurements or sets thereof are 



25 



WO 92/01928 



- 75 - 



PCT/US91/05071 



performed in the presence of said sample and calibrant 
fluids, respectively. 

8. The method of claim 4 or 5 in which said 
first fluid is a calibrant fluid. 

5 9 - The method of claim 8 in which said second 

fluid is sample fluid. 

10. The method of claim 4 or 5 in which said 
first fluid is sample fluid. 

11. The method of claim 10 in which said second 
10 fluid is calibrant. 

12. The method of claim l, 2, 3, 4 or 5 which 
further comprises providing a conductivity sensor capable 
of measuring the conductivity of a fluid in contact 
therewith . 

15 13 • The method of claim 1, 2 or 3 which further 

comprises performing a first conductivity measurement in 
the presence of said calibrant fluid. 

14. The method of claim 4 or 5 which further 
comprises performing a first conductivity measurement in 

20 the presence of said first fluid. 

15. The method of claim 13 which further 
comprises performing a second conductivity measurement in 
the presence of said sample fluid. 

16. The method of claim 14 which further 
25 comprises performing a second conductivity measurement in 

the presence of said second fluid. 



17. The method of claim 13 in which said first 
conductivity measurement is performed prior to performing 
said first signal measurement or set thereof. 



WO 92/01928 



- 76 - 



PCI7US91/05071 



10 



18. The method of claim 14 in which said first 
conductivity measurement is performed prior to performing 
said first signal measurement. 

19. The method of claim 15 in which said second 
conductivity measurement is performed after performing said 
second signal measurement or set thereof. 

20. The method of claim 16 in which said second 
conductivity measurement is performed after performing said 
second signal measurement. 

21. The method of claim 3 or 4 which further 
comprises activating said amperometric sensor or array 
thereof . 

22. The method of claim 21 in which activating 
said amperometric _ sensor comprises subjecting said 

15 amperometric sensor 'to a series of potential changes in the 
presence of said calibrant or first fluid. 

23. The method of claim 22 in which said 
potential. changes comprises cycling the applied potential 
between values of opposite sign relative to said reference 

20 electrode. 

24. The method of claim 22 in which said series 
of potential changes includes a first group which comprises 
at least ten repetitions of a cycle in which the applied 
potential is maintained at a first value and then is 

25 switched, to a second value of equal magnitude but of 
opposite sign as said first value. 

25. The method of claim 24 in which said series 
of potential changes further includes a second group which 
comprises at least five repetitions of a cycle in which the 

30 applied potential is maintained at a first value and then 



W ° 92/01928 PCT/US91/05071 

- 77 - 



10 



is switched to a second value of equal magnitude but of 
opposite sign as said first value. 

26. The method of claim 25 in which said series 
of potential changes further includes a third group which 
comprises at least five repetitions of a cycle in which the 
applied potential is maintained at a first value and then 
is switched to a second value of equal magnitude but of 
opposite sign as said first value. 

27. The method of claim 24 in which the 
application of said first group is commenced within about 
one second after said calibrant or first fluid comes into 
contact with said amperometric sensor. 

28. The method of claim 25 in which the 
application of said second group is commenced within about 

15 twelve seconds after said calibrant or first fluid comes 
into contact with said amperometric sensor. 

29. The method of claim 2 6 in which the 
application of said third group is commenced within about 
twenty seconds after said calibrant or first fluid comes 
into contact with said amperometric sensor. 



20 



30. The method of claim 25 in which said first 
value for said first and second groups is about +1000 mV. 

31. The method of claim 26 in which said first 
value for said third group is within double the magnitude 

25 of the operating potential of said amperometric sensor. 

32. The method of claim 3 which further 

comprises : 



providing a conductivity sensor capable of 
measuring the conductivity of a fluid in contact therewith, 



WO 92/01928 



- 78 - 



PCT/US91/05071 



activating said amperometric sensor in the 
presence of said calibrant fluid, and 

performing a first conductivity measurement in 
the presence of said calibrant fluid, after activating said 
5 amperometric sensor but before performing any set of signal 
measurements . 

33. The method of claim 32 which further 
comprises performing a second conductivity measurement in 
the presence of said sample fluid, after performing all 

10 sets of signal measurements. 

34. " The method of claim 1, 2 f 3, 4 or 5 in 
which said signal measurements or sets thereof are 
performed within about 2 minutes 

35. The method of claim 1, 2, 3, 4 or 5 in 
15 which the end of said first time window and the beginning 

of said second time window are separated by about three to 
about six seconds. 

.36. The method of claim 1, 2, 3, 4 or 5 in 
which the end of said first time window and the beginning 
20 of said second time window are separated by no more than 
about ten seconds. 

37. The method Of claim 1, 2, 3, 4 or 5 in 
which the duration of said time windows is about five to 
about fifteen seconds. 

25 38. The method of claim 3 in which one of said 

reference electrodes is replaced with a counter electrode, 

such that the remaining reference electrode is 
available to both of said potentiometric and amperometric 
sensors and said counter electrode is dedicated to said 
30 amperometric sensors • 



WO 92/01928 PCT/US91/05071 

- 79 - 



39. The method of claim 1, 2, or 3 in which the 
displacement of said calibrant fluid is carried out by the 
introduction of said sample fluid. 

40. The method of claim 39 in which the 
5 displacement of said calibrant fluid with said sample fluid 

is carried out with an air segment separating said fluids. 

41. The method of claim 4 or 5 in which the 
displacement of said first fluid is carried out by the 
introduction of said second fluid. 

10 42 ' The method of claim 41 in which the 

displacement of said first fluid with said second fluid is 
carried out with an air segment separating said fluids. 

43. The method of claim 1 in which performing 
said signal measurements comprises acquiring a preselected 

15 number of data points over the time period of said time 
windows . 

44. The method of claim 43 in which acquiring 
said data points comprises: 



20 



(a) collecting a first preselected fraction of 
said data points in a first sliding window, which first 
sliding window comprises a corresponding fraction of said 
time window; and 

(b) collecting additional preselected fractions 
of said data points in subsequent sliding windows until 

2.5 every data point in said time window is included in at 
least one sliding window. 



45. The method of claim 44 in which said first 
subsequent sliding windows are of equal duration. 



WO 92/01928 



- 80 - 



PCT/US91/05071 



46. The method of claim 44 in which said first 
and additional preselected fractions contain an equal 
number of data points . 

47. The method of claim 44 in which collecting 
5 additional fractions of said data points comprises 

including at least one new point of data in said first 
sliding window to form a subsequent sliding window . 

48. The method of claim 44 in which collecting 
additional fractions of said data points comprises removing 

10 at least the earliest point of data collected in said first 
sliding window and adding to the remaining points of data 
at least one new point of data to form a subsequent sliding 
window. 

49. The method of claim 44 which further 
15 comprises designating one of said preselected fractions as 

forming a basis set of data points and determining its 
characteristics . v 

50. The method of claim 4 9 which further 
comprises comparing the characteristics of said basis set 

20 of data points with the characteristics of the remaining 
preselected fractions. 

51. The method of claim 50 in which one of said 
characteristics includes the slope of each preselected 
fraction in its respective sliding window. 

25 52. The method of claim 50 which further 

comprises detecting the presence of aberrant data points on 
the basis of such comparisons. 

53. The method of claim 52 which further 
comprises counteracting the deleterious effects of said 
30 aberrant data points if the presence of such points has 
been detected. 



WO 92/01928 



- 81 - 



PCT/US91/05071 



54. The method of claim 53 in which 
counteracting the deleterious effects of said aberrant data 
points comprises performing a step selected from the group 
consisting of utilizing median filter techniques, 

5 eliminating the data points which lie outside the desired 
range and interpolating the remaining data points/ 
utilizing digital filter techniques or discarding the 
sliding window which contains said aberrant data points. 

55. The method of claim 1 which further 
10 comprises comparing the slope of the data points in a 

particular time window to a range of expected values to 
determine the integrity of the corresponding signal 
measurement. 

56. The method of claim 1 which further 
15 comprises comparing the difference between the slope of the 

data points in said .first time window and the slope of the 
data points in said second time window to a range of 
preselected limiting values to determine the integrity of 
said signal measurements. 

20 57. The method of claim 1 which further 

comprises comparing the mean value of the data points in a 
particular time window to a range of expected values to 
determine the integrity of said signal measurement. 

58. The method of claim 1 which further 
25 comprises comparing the deviation of the data points in a 
particular time window from the linear interpolation of 
said data points to a range of preselected limiting values 
to determine the integrity of said signal measurement. 



30 



59. The method of claim 1, 2, 3, 4 or 5 in 
which relating said signal measurements comprises: 



WO 92/01928 



- 82 - 



PCT/US91/05071 



(a) interpolating the data points in said first 
time window; 

(b) interpolating the data points in said 
second time window; 

5 (c) extrapolating said first time window 

interpolation forward to a position located between said 
first and second time windows; 

(d) extrapolating said second time window 
interpolation backward to said position; and 

10 (e) calculating the ratio of the values 

obtained from said extrapolations, from which ratio the 
concentration of said preselected analyte species can be 
determined. 

60. The -inethod of claims 1, 2 or 3 in which 
15 said calibrant fluid is comprised of an aqueous liquid or a 

wet gas . 

61. The method of claim 22 in which said 
potential * changes are applied in the form of pulses, 
incremental steps, sine waves, linear sweeps or 

20 combinations thereof. 

62. The method of claim l r 2, 3, 4 or 5 in 
which at least two signal measurements are performed in 
each of said first and second time windows, one of said 
signal measurements being carried out at a first applied 

25 potential and the other being carried out at a second 
applied potential. 

63. The method of claim 62 in which said signal 
measurement includes measuring the current. 



WO 92/01928 



- 83 - 



PCT/US91/05071 



10 



64. The method of claim 62 in which said first 
and second applied potentials lie in the range of about 100 
to about 300 mV. 

65. The method of claim 62 in which one of said 
signal measurements in said first time window is carried 
out at an applied potential of about 125 mV and the other 
measurement is carried out at about 225 mV. 

66. The method of claim 65 in which one of said 
signal measurements in said second time window is carried 
out at an applied potential of about 225 mV and the other 
measurement is carried out at about 125 mV. 



67. The method of claim 62 which further 
comprises determining the slope of the line defined by at 
least two points on a signal versus applied potential 

15 curve, which points are obtained from said signal 
measurements performed at each of said time windows. 

68. The method of claim 67 in which relating 
said signal measurements involves comparing the slopes of 
said lines' obtained from said time windows to determine the 

20 concentration of said preselected anaiyte species. 



WO 92/01928 



PCT/US91/05071 




WO 92/01928 



PCIYUS91/05071 




WO 92/01928 



PCT/US91/05071 



4/15 

o 




Millivolts 

O 

Ll. 



WO 92/01928 



PCT/US91/05071 




WO 92/01928 



PCT/US9I/05071 




WO 92/01928 



PCT/US91/05071 




WO 92/01928 



PCI7US91/05071 




WO 92/01928 



PCT/US91/05071 



FIG. 4 




WO 92/01928 



PCT/US91/05071 




WO 92/01928 



PCI7US91/0507! 




WO 92/01928 



PCT/US91/05071 




WO 92/01928 



PCT/US91/05071 




INTERNATIONAL SEARCH REPORT 

International Application No. PCT /DS91 /05071 



i. CLASSIFICATION OP 3U»JICT MATTtW (,l e.verai cla»„ncat,on »ymooi, apolyt , wcWcai » ttl) « 

^WW^Wi^^WW^ 31/00 G01F 1/66 
U.S.C1.: 204/400,402,406,153.1,424,433 73/861.28 



H FIILOS SEARCHED 



Clast>r. C j:.on S/stem 



U.S.C1. 
IPC (5) 



Minimum Doeum«nt«tion Searched » 



Cla*»ificat«on Symbols 



204/400,402,406,153.1,424,433 73/861.28 

COIN 27/30,27/54,27,00,31/12,27/02,27/46,27/28 G01P/1/66 



Documentation Searched other than Minimum Documentation 
to tho Extent that such Document! art Included in the Fields Searched * 



III. DOCUMENT* CONSIDERED TO BI RELEVANT • 



Cettaory 



X 
Y 
Y 
Y 
Y 
Y 
Y 
Y 



Citation of Document. " with indication, where appropriate, of the relevant passages 9 



US, A, 4,225,410 (PACE) 30 September 1980 * 
(See entire document), 

US, A, 4,935,106 (LISTQN ET AL.) 19 June 1990 
(See entire document). 

US, A, 4,935,105 (GHURGHHOUSE) 19 June 1990 
(See entire document). 

US, A, 4,686,479 (YOUNG EI AL.) 11 August 1987 
(See entire document). 

US, A, 4,293,522 (WINKLER) 06 December 1984 
(See entire docunent). 

US, A, 4,033,830 (FLETCHER, III) 05 July 1977 
(See entire document). 

US, A, 4,535,285 (EVANS ET AL.) 13 August 1985 
(See entire document). 

US, A, 4,019,966 (REMES ET AL.) 26 April 1977 
(See entire document). 



Relevent to Claim No. * 



1-7, 9-11 
1-12, 15, 16 
1-5, 21,22,34 
10-18, 60 
19, 20 

22- 26, 38 

23- 29,32,34-37 
38 



* Special ca tegories of cited documents: « 
"A" document defining tho oonerei state of the art which la not 
considered to be) of particular relevance 

"V earlier document but pubtJehed on or after tho international 
filing data 

document which may threw doubts on priority doimU) er 
which if cited to estebiishthe puhhcation date of another 
citation or other apecia! reeeon (at opacified) 
-O* document referring to an oral disclosure, use, exhibition er 



"V later document published after tho International flUng date 
or priority date and not hi conflict with the application but 
cited to understand the principle or theory underlying tho 
invention 

M X" document of particufar relevance; the claimed invention 
cannot be considered novel or cannot be considered to 



T* document of parpcuiar _ . . _ 
cannot be considered to invoke 
lit combined with 



»P- document published prior to the intarnational filing oat* but 
later than the priority date claimed 



in tho art 
"4* document m ember of the 



... the 

an mventrie step when the 
> or more other such doce- 
obvioos to « person skated 



iv. ctirrtncATiON 



Date of the Actual Completion of 

16 October 1991 



the International Search 



Oats of Mailing of this International Search Report 

25 NOV 1991 / 




ol Auttwtttad Oflfetr 



ISA/US 



International Application No/ pgj /US91/05071 



FURTHER INFORMATION CONTINUE* FROM THE SECOND SHEET 



A 



US, A, 3,672,843 (ROSSE EI AL.) 27 June 1972 
(Sea entire document). 

US, A, 4,787,252 (JACOSSON ET AL. ) 29 November 1988 
(See entire docisnent). 

US, A, 4,464,230 (UHK3)0N) 07 August 1984 
(See entire document) . 

US, A, 4,897,162 (LEWANDOWSKI ET AL.) 30 January 
( See entire document ) . 1990 



39, 41 
43-59,62 
61, 63 



OBSERVATIONS WHERE CERTAIN CLAIMS WERE FOUND UNSEARCHABLE 1 



This international search report has not botn established la respect of certain claims under Article 17(2) (a) for the following reasons: 
t. □ Claim numbers . because they relata to subject matter M not required to be searched by this Authority, namely: 



2-0 C 1 *'" 1 numbers ... 

ments to such an extent 



because they relate to parts of the international application that do not comply with the prescribed require- 
that no meaningful international search can be carried out », specifically: 



xQ ClaJm numbers^ 



, be*«e^sy»depe*de^ 



PCT Rule 6.4(a). 



VI. Q OBSERVATIONS WHERE UNITY OW INVENTION IS LACKING » 



This Internationai Sesrchlno Authority found multipJo inventions In this International application as follows: 



.1.Q As all required additional search fees were timely paid by the applicant, this international search report covers all searchable claims 
of the international application. ' 

MD As only soma of the required additional search fees were timer, paid by the applicant this international search report covers only 
those claims of the Intsmstionsl application for which fees wars paid, spectflcsfly dalmss 



3.Q No required additional search fees were time* paid by the applicant Consequently, this International search report Is restricted 1 
the invention first mentioned in the claims; a Is covered by claim numbers: 



be searched without effort fustifying an additional left, the International Searching Authority did not 



4.n As aUsearebableclaims could 

UJ invite payment of any additional fee. 



Q The additional search lees were 
□ No protest accompanied the payment of 
_______ 



accompanied by applicant's protest 



I "»»™««on»l Application N».~PCT7US91 /05071 

J ^DOCUWINTS COH8.DIW.P TO £ HEL.V ANT (COWTIHU1D ««0, THE SECOND SHIT, 



C *" g ° ry ' I """O" »' Pocunwnt, ■» with IndlcUon, wh»r« .pprapriw,. of »» rrtwinl 



US, A, 4^86,590 (TITTLE) 12 DECEMBER 1989 
(See entire document). 

US, A, 3,765,841 (PAULSEN ET AL.) 

16 OCTOBER 1973 (See entire document). 



R»!*r»nt to Claim No » 



PCT/tSA/210 (extra *h*tt) (May 19S8) 



THIS PAGE BLANK (uspto)